Operating information
The bottom portion of the window provides more details about the instrument oper-
ating history and status.
l
Pump Operating Hours
is the number of hours that the pump has operated.
l
Thermal Enclosure Temp
is the temperature of components inside the foam insu-
lating enclosure. If red, the instrument is either too hot and unable to regulate
the enclosure temperature, or too cold and in a warmup state. Normally this is
between 51.5 °C and 57.5 °C.
l
Cavity Temperature
is the optical bench temperature. If red, the instrument is
either too hot and unable to regulate the cavity temperature, or too cold and in a
warmup state. Normally this is between 52 °C and 58 °C.
l
Laser Temperature
is the temperature of the diode laser. If red, the instrument is
unable to regulate the laser temperature.
l
Laser Temp. Shift
is the shift in the laser center wavelength from the factory cal-
ibration.
l
Ring Down Time
is a parameter indicating cavity resonance.
l
Device Uptime
shows the number of hours that the instrument has been
powered on.
l
On/Off cycles
is the number of times the instrument has been powered on.
Section 4. Software overview
4-13
Operating information
Содержание LI-7815
Страница 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 10: ...viii...
Страница 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Страница 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Страница 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Страница 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Страница 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Страница 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Страница 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Страница 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Страница 97: ......