Section 7. Communication with the MQTT protocol
Rebooting the instrument
To restart the instrument, send the following command:
mosquitto_pub -h
172.24.82.41
-t
"licor/niobrara/system/reboot"
-m
"reboot"
This initiates the restart cycle, beginning with a purge of the optics.
Powering off the instrument
To power off the instrument, send the following command:
mosquitto_pub -h
172.24.82.41
-t
"licor/niobrara/system/shutdown"
-m
"shutdown"
This initiates the shutdown, beginning with a purge of the optics. There is no com-
mand to power on the instrument remotely.
7-6
LI-7815 CO₂/H₂O Gas Analyzer