Section 6. Maintenance
Setting the CO₂ zero
If you want to measure absolute CO
2
concentrations, in soil efflux applications, for
example, you can set the CO
2
zero. This is supported in software versions 2.0.17 and
newer. You can adjust the zero with a CO
2
-free gas—either tank gas that is free of
volatile organic compounds (VOCs) or air that has been scrubbed of CO
2
using the
CO
2
Zero Kit (7800-950). To zero CO
2
, plumb tank gas to the air inlet or use the
zero kit. When the CO
2
reading is stable, click the
CO
2
Zero
button. If you are unsat-
isfied with the zero operation, click
Clear
to restore the most recent instrument zero
(either the factory zero setting or the settings applied with a general zero calibration.
Recovering from a bad calibration
In the event that an attempt to zero or span CH
4
fails to work as intended, you can
use previous calibration data under
Restore Span/Zero
. Select the file and click
Restore
.
Factory Reset
reloads the factory original zero and span calibration parameters.
6-4
LI-7810 CH₄/CO₂/H₂O Trace Gas Analyzer
Содержание LI-7810
Страница 1: ...LI 7810 CH CO H O Trace Gas Analyzer Instruction Manual ...
Страница 2: ......
Страница 6: ...iv ...
Страница 30: ...Section 2 Getting started 2 14 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 38: ...Section 3 Applications of the LI 7810 3 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 54: ...Section 4 Software overview 4 16 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 62: ...Section 5 Troubleshooting 5 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 88: ...Appendix B Additional safety information B 6 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 92: ...Appendix C Specifications C 4 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 97: ......