Section 6. Maintenance
Setting the CH₄, CO₂, and H₂O zeros
Select a quality zero gas that is free of volatile organic compounds (VOCs). The cal-
ibration will only be as good as the gas. The gas should be 0 ppm CH
4
and 0 ppm
CO
2
in air.
Caution:
Do not use nitrogen-balanced gas. Use air-balanced gas for both the zero
and span. Do not power on the instrument with a 0 ppm methane gas flowing
through the analyzer or the analyzer will not initialize properly.
1
Power on the instrument and let it warm up.
2
Plumb a gas to the air inlet as shown in
Set the pressure between ambient and 35 kPa (5 PSI) above ambient.
Observe the measurements on the home screen until they have stabilized. The slope
will be near zero.
3
Zero the analyzer.
Under
Options > Calibration
, click
Zero
. The software will indicate that a zero cal-
ibration has started and it will not report concentrations until the zero is set. This
usually takes about 5 minutes.
6-2
LI-7810 CH₄/CO₂/H₂O Trace Gas Analyzer
Содержание LI-7810
Страница 1: ...LI 7810 CH CO H O Trace Gas Analyzer Instruction Manual ...
Страница 2: ......
Страница 6: ...iv ...
Страница 30: ...Section 2 Getting started 2 14 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 38: ...Section 3 Applications of the LI 7810 3 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 54: ...Section 4 Software overview 4 16 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 62: ...Section 5 Troubleshooting 5 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 88: ...Appendix B Additional safety information B 6 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 92: ...Appendix C Specifications C 4 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 97: ......