Status and warning messages
During warmup and any time the instrument detects a problem, it will issue one or
more status messages, including a message that says SVC Required. This is normal
during start-up. Typically, start-up messages will be resolved within an hour of oper-
ation. The status messages will be triggered again if there is an issue with the system.
See
on page 2-9 for details on the codes.
Code 128 triggered
Code 128 is triggered if the inlet is obstructed. This can happen if the instrument
has drawn water, dust, or debris into the inlet filter, causing a clog. In any case,
power off the instrument, clear the obstruction, replace the filters (see
on page 6-5), and power the instrument back on.
You must restart the instrument to resolve code 128.
Section 5. Troubleshooting
5-7
Status and warning messages
Содержание LI-7810
Страница 1: ...LI 7810 CH CO H O Trace Gas Analyzer Instruction Manual ...
Страница 2: ......
Страница 6: ...iv ...
Страница 30: ...Section 2 Getting started 2 14 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 38: ...Section 3 Applications of the LI 7810 3 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 54: ...Section 4 Software overview 4 16 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 62: ...Section 5 Troubleshooting 5 8 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 88: ...Appendix B Additional safety information B 6 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 92: ...Appendix C Specifications C 4 LI 7810 CH CO H O Trace Gas Analyzer ...
Страница 97: ......