Section 7.
Maintenance
In this section we describe maintenance procedures.
Changing the fluorescent tube
l
Grasp the tube near each end by the edges of the upper transparent belt. Twist
the tube 1/4 revolution to loosen it in the connectors.
l
Remove the rear tube connector by loosening the two recessed mounting screws.
l
Draw the tube out through the hole in the instrument rear plate.
l
Install the replacement tube as described in
Changing the transparent belts
The following steps are used to remove the transparent belts. An alternative tech-
nique is outlined in
Alternative belt removal technique
1
Disconnect the power.
2
Remove the rear housing (
3
Loosen the transparent belts (
Transparent belt tension adjustment
4
Remove the fluorescent tube (
5
Separate the fluorescent tube power cord connector located near the instrument front plate and
under the belt drive pulleys.
This connector may remain mated if the front plate can rest near the instrument
without straining the cord.
7-1
LI-3100C Area Meter Manual
Содержание LI-3100C
Страница 1: ...LI 3100C Area Meter Instruction Manual...
Страница 2: ......
Страница 8: ...Section 1 General information 1 2 LI 3100C Area Meter Manual...
Страница 14: ...Section 2 Preparations 2 6 LI 3100C Area Meter Manual...
Страница 19: ...Figure 3 2 Press roller in position for use with thin leaves Section 3 Operation 3 5 Pressing roller adjustment...
Страница 20: ...Section 3 Operation 3 6 LI 3100C Area Meter Manual...
Страница 24: ...Section 4 Changing measurement resolution 4 4 LI 3100C Area Meter Manual...
Страница 32: ...Section 5 Using the Windows interface software 5 8 LI 3100C Area Meter Manual...
Страница 36: ...Section 6 Troubleshooting 6 4 LI 3100C Area Meter Manual...
Страница 41: ...brush Section 7 Maintenance 7 5 Cleaning the motor and belt drive system...
Страница 42: ...Section 7 Maintenance 7 6 LI 3100C Area Meter Manual...
Страница 49: ......