Changing from 0.1 mm² to 1.0 mm² resolution
The procedures for resolution change are outlined in the following steps.
1
Disconnect the power.
2
Remove the rear housing by removing the four retaining screws.
3
Move the decimal selector switch to the 1.0 mm
2
position.
4
Loosen all four set screws on the outer camera pressure rail.
Figure 4-2. LI-3100C camera position configured for 1.0 mm
2
operation.
5
Align the yellow focus mark to the red lens index mark.
6
Lightly tighten the set screws.
7
Be sure the lens cover has been removed and that no other obstruction is in the optical path.
8
Replace the rear housing.
Calibration should not be necessary; check the calibration if desired.
Section 4. Changing measurement resolution
4-3
Changing from 0.1 mm² to 1.0 mm² resolution
Содержание LI-3100C
Страница 1: ...LI 3100C Area Meter Instruction Manual...
Страница 2: ......
Страница 8: ...Section 1 General information 1 2 LI 3100C Area Meter Manual...
Страница 14: ...Section 2 Preparations 2 6 LI 3100C Area Meter Manual...
Страница 19: ...Figure 3 2 Press roller in position for use with thin leaves Section 3 Operation 3 5 Pressing roller adjustment...
Страница 20: ...Section 3 Operation 3 6 LI 3100C Area Meter Manual...
Страница 24: ...Section 4 Changing measurement resolution 4 4 LI 3100C Area Meter Manual...
Страница 32: ...Section 5 Using the Windows interface software 5 8 LI 3100C Area Meter Manual...
Страница 36: ...Section 6 Troubleshooting 6 4 LI 3100C Area Meter Manual...
Страница 41: ...brush Section 7 Maintenance 7 5 Cleaning the motor and belt drive system...
Страница 42: ...Section 7 Maintenance 7 6 LI 3100C Area Meter Manual...
Страница 49: ......