Section 4.
Changing measurement resolution
The LI-3100C utilizes the full measurement width at all times, for both 0.1 and 1.0
mm
2
resolution. The 0.1 mm
2
resolution may be preferable , since it provides the
best accuracy, particularly with small or complex shaped objects to 0.5 mm in width.
The 1.0 mm
2
resolution may be preferable where large samples are being measured,
and the continuous buildup of debris on the belts causes excessive spurious counts.
Changing from 1.0 mm² to 0.1 mm² resolution
The procedures for the resolution change are outlined in the following steps.
1
Disconnect the power.
2
Remove the rear housing by removing the four retaining screws.
3
Move the decimal selector switch to the 0.1 mm
2
position (labeled near the switch). This switch is
located at the top right corner (as viewed from the rear of the display circuit board.
4
Loosen all four set screws on the outer camera pressure rail.
5
Align the red focus mark to the red lens index mark on the inner camera alignment rail.
4-1
LI-3100C Area Meter Manual
Содержание LI-3100C
Страница 1: ...LI 3100C Area Meter Instruction Manual...
Страница 2: ......
Страница 8: ...Section 1 General information 1 2 LI 3100C Area Meter Manual...
Страница 14: ...Section 2 Preparations 2 6 LI 3100C Area Meter Manual...
Страница 19: ...Figure 3 2 Press roller in position for use with thin leaves Section 3 Operation 3 5 Pressing roller adjustment...
Страница 20: ...Section 3 Operation 3 6 LI 3100C Area Meter Manual...
Страница 24: ...Section 4 Changing measurement resolution 4 4 LI 3100C Area Meter Manual...
Страница 32: ...Section 5 Using the Windows interface software 5 8 LI 3100C Area Meter Manual...
Страница 36: ...Section 6 Troubleshooting 6 4 LI 3100C Area Meter Manual...
Страница 41: ...brush Section 7 Maintenance 7 5 Cleaning the motor and belt drive system...
Страница 42: ...Section 7 Maintenance 7 6 LI 3100C Area Meter Manual...
Страница 49: ......