Installation
37
300855170_002_C4 - 06/2020 - © Leybold
Fig. 3.13 REMOTE interface X1
Pin Name
Description
1
Error relay
(com)
2
Error relay
(n.c.)
3
Normal operation relay
(n.o.)
4
Normal operation relay
(com)
5
Normal operation relay
(n.c.)
6
Signal GND
7
High level output
24 V, 100 mA, Tolerance acc. to
device supply voltage
8
Start input (High)
High > 10 V ± 0,5 V
Reset input (Low)
Low < 7.5 V ± 0,5 V
9
Error relay
(n.o.)
10
Standby input
High > 10 V ± 0.5 V
Low < 7.5 V ± 0.5 V
11
Cooling/Venting valve
High > 10 V ± 0.5 V
input (Low)
Low < 7.5 V ± 0.5 V
12
Analog output (Default: Frequency)
0 ... 10 V, 2 mA
13
Analog GND
14
Warning relay
(n.c.)
15
Warning relay
(com)
Shield Connected with pump housing
24 V
7
8
6
0 V = STOP/Fehlerreset
24 V = START
24 V
7
8
6
24 V
5 V
1k8Ω
3k3Ω
7
8
6
Steuerung
Kontakt offen = STOP//Fehlerreset
Kontakt geschlossen = START
5
3
4
n. c.
n. o.
com.
2
9
1
n. c.
n. o.
com.
Frequenzwandler
Frequenzwandler
14
15
n. c.
com.
11
5
1
15
6
10
REMOTE
X1
15-pole Sub-D female
High Density
24 V
7
8
6
0 V = STOP/Fehlerreset
24 V = START
24 V
7
8
6
24 V
5 V
1k8Ω
3k3Ω
7
8
6
Steuerung
Kontakt offen = STOP//Fehlerreset
Kontakt geschlossen = START
5
3
4
n. c.
n. o.
com.
2
9
1
n. c.
n. o.
com.
Frequenz-
wandler
Frequenz-
wandler
14
15
n. c.
com.
11
5
1
15
6
10
12
13
Analog-Ausgang
Steuerung
Frequenz-
wandler
Start/Stop input
Pin 10
0 V = no Standby operation
24 V = Standby operation
Pin 11
0 V = no function
24 V = Cooling or valve is activated
Pin 10
Contact open: no Standby operation
Contact closed: Standby operation
Pin 11
Contact open: no function
Contact closed: Cooling or valve is activated
24 V
7
8
6
0 V = STOP/Fehlerreset
24 V = START
24 V
7
8
6
24 V
5 V
1k8Ω
3k3Ω
7
8
6
Steuerung
Kontakt offen = STOP//Fehlerreset
Kontakt geschlossen = START
5
3
4
n. c.
n. o.
com.
2
9
1
n. c.
n. o.
com.
Frequenzwandler
Frequenzwandler
14
15
n. c.
com.
11
5
1
15
6
10
Relay
Normal operation relay
4 and 5 connected (normal position)
Run-down, run-up, stop
4 and 3 connected
Normal operation
Error relay
1 and 2 connected (normal position)
No error
1 and 9 connected
Error
Warning relay
14 and 15 connected (normal position) No warning
14 and 15 open
Warning
Analog output
Maintain pin 6 Signal
GND and pin 13 Analog
GND separate so as to
avoid equalisation cur-
rents.
Contact open = STOP/Error reset
Contact closed = START
0V = STOP/Error reset
24 V = START
Frequency
converter
Controller
Frequency
converter
Controller
Frequency
converter