16
GA300553102_002_C2
– 04/2017
6.
Troubleshooting
6.1
Indication of Errors
A malfunction in the GRAPHIX Controller is displayed by an error message on the screen
or issued by an error number via the serial interface. Additionally, an entry is made into the
error memory list, from which the 20 most recently registered errors can be displayed on the
screen (
Chapter 6.2
Error Log, page 17) as well as read out via serial interface. The
following tables give a survey on all recognisable sensor errors (
device errors (
6.1.1
Sensor Errors
Error description
(Error Log)
Display
(TFT display)
Error number
(Interface)
Cause and remedy
no error
Measuring
value
0
Attached sensor is recognized and
in the specified measuring range.
Sensor-ID short circuit
….
1
Error in the electric circuit of sensor
identification.
Sensor-ID unknown or not
available
….
2
Identification resistor of sensor
identification unknown or missing.
No sensor signal
Error-S
1
3
Fault affecting the connection to the
sensor. The message will only be
displayed in the display field of the
affected measurement channel.
Acknowledge this message by
selecting
and
terminating
the
channel menu at the corresponding
channel.
Analog sensor signal out of range -
too high
Error-H
1
4
The measurement signal from the
sensor is significantly above the
permissible range.
Analog sensor signal out of range -
too low
Error-L
1
5
The measurement signal from the
sensor is significantly below the
permissible range.
Communication error digital sensor
signal
Error-00
1
6
Communication error. Failure in the
data transfer to IONIVAC sensors of
ITR90, ITR200 and CTR-N series.
Error electronic/eeprom
Error-40
1
7
Electronics/EEPROM error with
IONIVAC sensors of ITR200 series.
Error pirani
Error-04
2
8
Pirani error with IONIVAC sensors
of ITR90 series.
Both filaments broken
Error-10
1
9
Hot cathode error (both filaments
defectively) with IONIVAC sensors
of ITR200 series.
One filament broken
Error-20
2
10
Hot cathode error (filament 1
defectively) with IONIVAC sensors
of ITR200 series.
Error pirani
Error-90
1
11
Pirani error with IONIVAC sensors
of ITR90 series.
Pirani adjustment out of range
Error-50
2
12
Pirani adjustment deficient with
IONIVAC sensors of ITR90 series.
Error ion gauge
Error-80
1
13
Hot
cathode
error
(filament
defectively) with IONIVAC sensors
of ITR90 series.
Table 7
–
Sensor errors
Содержание 230680V01
Страница 2: ......