background image

16 

 

GA300553102_002_C2 

– 04/2017 

 

6. 

Troubleshooting 

6.1 

Indication of Errors 

A malfunction in the GRAPHIX Controller is displayed by an error message on the  screen 
or issued by an error number via the serial interface. Additionally, an entry is made into the 
error memory list, from which the 20 most recently registered errors can be displayed on the 
screen (



 

Chapter 6.2

 

Error Log, page 17) as well as read out via serial interface. The 

following tables give a survey on all recognisable sensor errors (



 

Table 7, page 16) and 

device errors (



 

Table 8, page 17). 

 

6.1.1 

Sensor Errors 

Error description 

(Error Log) 

Display 

(TFT display) 

Error number  

(Interface) 

Cause and remedy 

no error 

Measuring 

value 

Attached sensor  is recognized  and 
in the specified measuring range. 

Sensor-ID short circuit 

…. 

Error in the electric circuit of sensor 
identification. 

Sensor-ID unknown or not 
available 

…. 

Identification  resistor  of  sensor 
identification unknown or missing. 

No sensor signal 

 

Error-S

 1 

Fault affecting the connection to the 
sensor.  The  message  will  only  be 
displayed in the display field of the 
affected  measurement  channel. 
Acknowledge  this  message  by 
selecting 

and 

terminating 

the 

channel menu at the corresponding 
channel.

 

Analog sensor signal out of range - 
too high 

 

Error-H

 1 

The measurement signal from the 
sensor is significantly above the 
permissible range.  

Analog sensor signal out of range - 
too low 

 

Error-L

 1

 

The  measurement  signal  from  the 
sensor  is  significantly  below  the 
permissible range.

 

Communication error digital sensor 
signal 

 

Error-00

 1

 

Communication error. Failure in the 
data transfer to IONIVAC sensors of 
ITR90, ITR200 and CTR-N series.

 

Error electronic/eeprom 

 

Error-40

 1

 

Electronics/EEPROM  error  with 
IONIVAC sensors of ITR200 series.

 

Error pirani 

 

Error-04

 2

 

Pirani  error  with  IONIVAC  sensors 
of ITR90 series.

 

Both filaments broken 

 

Error-10

 1

 

Hot  cathode  error  (both  filaments 
defectively)  with  IONIVAC  sensors 
of ITR200 series.

 

One filament broken 

 

Error-20

 2

 

10 

Hot  cathode  error  (filament  1 
defectively)  with  IONIVAC  sensors 
of ITR200 series.

 

Error pirani 

 

Error-90

 1

 

11 

Pirani  error  with  IONIVAC  sensors 
of ITR90 series.

 

Pirani adjustment out of range 

 

Error-50

 2

 

12 

Pirani  adjustment  deficient  with 
IONIVAC sensors of ITR90 series.

 

Error ion gauge 

 

Error-80

 1

 

13 

Hot 

cathode 

error 

(filament 

defectively)  with  IONIVAC  sensors 
of ITR90 series.

 

 

Table 7 

– 

Sensor errors 

 
 
 
 

Содержание 230680V01

Страница 1: ...Catalog No 230680V01 230681V01 230682V01 GRAPHIXONE GRAPHIXTWO GRAPHIXTHREE Vacuum Gauge Controller Operation Manual GA300553102_002_C2...

Страница 2: ......

Страница 3: ...Symbols 10 4 4 1 Symbols for the Controls 10 4 4 2 Symbols for the Language Selection 11 4 4 3 Status and Warning Symbols 11 4 5 Measuring Mode 12 5 Maintenance and Service 13 5 1 Maintenance 13 5 2 C...

Страница 4: ...HIX controller is a display and operating unit with a graphic user interface for sensors with an analogue or digital interface of the company Leybold GmbH or from other manufacturers Depending on the...

Страница 5: ...sume any responsibility or warranty in case the operator or third persons do not observe the information given in this document do not use the product as intended modify the product in any way convers...

Страница 6: ...h electricity NOTICE General information pointing to further information respectively reference sections 2 3 Basic Safety Information During all work like installation maintenance and repair activitie...

Страница 7: ...measurement channel 2 for GRAPHIX TWO and THREE only for sensors CH2 A and CH2 B E1 E2 Connection of measurement channel 3 for GRAPHIX THREE only for sensors CH3 A and CH3 B F Relay output connection...

Страница 8: ...sible You may select between the following display modes Normal Default display mode which provides all important information on the connected sensors Chart Graphic display of the pressure history of...

Страница 9: ...e functions In this way you may configure your GRAPHIX controller and also utilise further functions offered by the instrument Channel Menu Buttons Briefly touching the desired channel invokes the cha...

Страница 10: ...Enabling display mode Leak Test Speedo Channel 1 Enabling display mode Speedo for channel 1 Speedo Channel 2 Enabling display mode Speedo for channel 2 Speedo Channel 3 Enabling display mode Speedo f...

Страница 11: ...ge RU Russian Polish Select menu language PL Polish Turkish Select menu language TR Turkish Table 4 Symbols for language selection 4 4 3 Status and Warning Symbols Symbol Designation Explanation Statu...

Страница 12: ...in specified measuring range Measuring value and additional warning symbol in the display mode Normal and Speedo or Measuring value displayed by a yellow font in the display modes Chart Big and Leak...

Страница 13: ...n the USB stick in the folder RECOVERY o After the reset process has been completed the instrument will automatically restart with the default parameters o The GRAPHIX controller will now be ready for...

Страница 14: ...Setup o The question Are you sure is displayed to ensure that you really want to run the reset When not wanting to run the reset tap the button to cancel the reset process o The main menu is then disp...

Страница 15: ...yed again To run the update tap the button Start update o On the memory at the USB interface is installed the folder RECOVERY o Configuration data are stored in a file with the format rescue txt on th...

Страница 16: ...measurement channel Acknowledge this message by selecting and terminating the channel menu at the corresponding channel Analog sensor signal out of range too high Error H 1 4 The measurement signal f...

Страница 17: ...memory stick Error usb file closing none 3 File on USB stick cannot be stored Check the attached USB memory stick Error usb write none 4 USB stick not recordably Check the attached USB memory stick Er...

Страница 18: ...n The setup value is saved To exit tap the button o The parameter selection display is displayed again 6 3 Help in Case of Malfunctions If the malfunction persists even after having exchanged the sens...

Страница 19: ...ry room During storage the following ambient conditions need to be maintained Ambient temperature 20 60 C Humidity of the air As low as possible Preferably in a sealed plastic bag with desiccant 7 3 W...

Страница 20: ...rvice zh leybold com Spain Leybold Spain S A C Huelva 7 E 08940 Cornell de Llobregat Barcelona Sales T 34 93 666 43 11 F 34 93 666 43 70 sales ba leybold com Service T 34 93 666 46 11 F 34 93 685 43 7...

Отзывы: