12
GA300553102_002_C2
– 04/2017
4.5
Measuring Mode
The measuring mode is the standard mode of operation. Here the measured values of the
sensors in the display modes Normal, Big or Chart are shown. Additionally status messages
(
Table 6, page 12) and/or error messages (
Table 7, page 16) can be displayed.
Display
Description
....
No sensor connected.
FS?
Full Scale? Connected analogue CERAVAC sensor
has not been specified. Make a selection, in order to
specify the sensor.
S-OFF
High
vacuum
measurement
circuit
of
the
PENNINGVAC sensors PTR81N, PTR225, TR225N,
PTR225S, PTR225SN, PTR237 or PTR237N has
been switched off.
Measuring value
Connected sensor is identified and in specified
measuring range.
Measuring value
and additional warning symbol
in the display mode Normal and Speedo
or
Measuring value
displayed by a
yellow font
in the
display modes Chart, Big and Leak Test as well as
for the other channels in the display mode Speedo
Description depends on the connected sensor:
•
Pirani adjustment of the connected IONIVAC
sensor of ITR90 series is insufficient.
•
Filament 1 of the connected IONIVAC sensor of
ITR200 series is defective.
•
Connected CERAVAC sensor of CTR101 series is
in the heating phase.
Table 6
–
Status messages measured values display
After switching on, the GRAPHIX controller will automatically resume the last measurement
display mode, which was selected. When running the main menu and not making an entry
for more than 60 seconds, then the instrument will revert back to the measured values
display.