background image

GA300553102_002_C2 

– 04/2017

 

 

17

 

 

1

 In addition to display of the error in place of the measured value, the warning symbol 

 blinks in display 

modes Normal or Speedo. The error message is displayed in red font for the display modes Chart, Big and 
Leak Test as well as for the other channels in display mode Speedo. The error signal relay toggles according 
to the settings. 

2

 In addition to the measured value, the warning symbol 

 glows in display modes Normal or Speedo. The 

error message is displayed in yellow font for the display modes Chart, Big and Leak Test as well as for the 
other channels in display mode Speedo. 

 

6.1.2 

System Errors 

Error description 

(Error Log) 

Display 

(TFT display) 

Error number  

(Interface) 

Cause and remedy 

no errror 

none 

System works error-free. 

Error usb file open 

none

 

File  on  USB  stick  cannot  be 
opened.  Check  the  attached  USB 
memory stick. 

Error usb folder open 

none

 

Folder  on  USB  stick  cannot  be 
opened.  Check  the  attached  USB 
memory stick.

 

Error usb file closing 

none

 

File on USB stick cannot be stored. 
Check  the  attached  USB  memory 
stick. 

Error usb write 

none

 

USB  stick  not  recordably.  Check 
the attached USB memory stick. 

Error usb read 

none

 

USB stick not readably.  Check the 
attached USB memory stick. 

Error read eeprom 

none

 

EEPROM  of  the  equipment  not 
readably. 

Contact 

your 

next 

Leybold GmbH service office. 

Error write eeprom 

none

 

EEPROM of system not recordably. 
Contact  your  next  Leybold  GmbH 
service office. 

Error init eeprom 

none

 

EEPROM error of system. 
Contact  your  next  Leybold  GmbH 
service office. 

 

Table 8 

– 

System errors

 

 

6.2 

Error Log 

The  GRAPHIX  Controller  stores  the  errors  recognized  by  the  system.  They  have  the 
possibility of being able to be been noticeable the last 20 errors. Important information on 
the display area is represented

 

(



 Fig. 5, page 17)

 

 

 

 

 

Fig. 5 

 Error log 

 
A

 

Error number 1 

– 20 

 

(1 = freshest error) 

Date of displayed error 

Time of displayed error 

Error source: 

 

Channel 1 = Sensor error at channel 1 

 

Channel 2 = Sensor error at channel 2 

 

Channel 3 = Sensor error at channel 3 

 

Device = System error 

Error description 

 

(



 

Table 7, page 16 and Table 8, 

 

page 17) 

Содержание 230680V01

Страница 1: ...Catalog No 230680V01 230681V01 230682V01 GRAPHIXONE GRAPHIXTWO GRAPHIXTHREE Vacuum Gauge Controller Operation Manual GA300553102_002_C2...

Страница 2: ......

Страница 3: ...Symbols 10 4 4 1 Symbols for the Controls 10 4 4 2 Symbols for the Language Selection 11 4 4 3 Status and Warning Symbols 11 4 5 Measuring Mode 12 5 Maintenance and Service 13 5 1 Maintenance 13 5 2 C...

Страница 4: ...HIX controller is a display and operating unit with a graphic user interface for sensors with an analogue or digital interface of the company Leybold GmbH or from other manufacturers Depending on the...

Страница 5: ...sume any responsibility or warranty in case the operator or third persons do not observe the information given in this document do not use the product as intended modify the product in any way convers...

Страница 6: ...h electricity NOTICE General information pointing to further information respectively reference sections 2 3 Basic Safety Information During all work like installation maintenance and repair activitie...

Страница 7: ...measurement channel 2 for GRAPHIX TWO and THREE only for sensors CH2 A and CH2 B E1 E2 Connection of measurement channel 3 for GRAPHIX THREE only for sensors CH3 A and CH3 B F Relay output connection...

Страница 8: ...sible You may select between the following display modes Normal Default display mode which provides all important information on the connected sensors Chart Graphic display of the pressure history of...

Страница 9: ...e functions In this way you may configure your GRAPHIX controller and also utilise further functions offered by the instrument Channel Menu Buttons Briefly touching the desired channel invokes the cha...

Страница 10: ...Enabling display mode Leak Test Speedo Channel 1 Enabling display mode Speedo for channel 1 Speedo Channel 2 Enabling display mode Speedo for channel 2 Speedo Channel 3 Enabling display mode Speedo f...

Страница 11: ...ge RU Russian Polish Select menu language PL Polish Turkish Select menu language TR Turkish Table 4 Symbols for language selection 4 4 3 Status and Warning Symbols Symbol Designation Explanation Statu...

Страница 12: ...in specified measuring range Measuring value and additional warning symbol in the display mode Normal and Speedo or Measuring value displayed by a yellow font in the display modes Chart Big and Leak...

Страница 13: ...n the USB stick in the folder RECOVERY o After the reset process has been completed the instrument will automatically restart with the default parameters o The GRAPHIX controller will now be ready for...

Страница 14: ...Setup o The question Are you sure is displayed to ensure that you really want to run the reset When not wanting to run the reset tap the button to cancel the reset process o The main menu is then disp...

Страница 15: ...yed again To run the update tap the button Start update o On the memory at the USB interface is installed the folder RECOVERY o Configuration data are stored in a file with the format rescue txt on th...

Страница 16: ...measurement channel Acknowledge this message by selecting and terminating the channel menu at the corresponding channel Analog sensor signal out of range too high Error H 1 4 The measurement signal f...

Страница 17: ...memory stick Error usb file closing none 3 File on USB stick cannot be stored Check the attached USB memory stick Error usb write none 4 USB stick not recordably Check the attached USB memory stick Er...

Страница 18: ...n The setup value is saved To exit tap the button o The parameter selection display is displayed again 6 3 Help in Case of Malfunctions If the malfunction persists even after having exchanged the sens...

Страница 19: ...ry room During storage the following ambient conditions need to be maintained Ambient temperature 20 60 C Humidity of the air As low as possible Preferably in a sealed plastic bag with desiccant 7 3 W...

Страница 20: ...rvice zh leybold com Spain Leybold Spain S A C Huelva 7 E 08940 Cornell de Llobregat Barcelona Sales T 34 93 666 43 11 F 34 93 666 43 70 sales ba leybold com Service T 34 93 666 46 11 F 34 93 685 43 7...

Отзывы: