18
ENGLISH
Immunity test
Test level IEC60601
CONFORMITY
level
ELECTROMAGNETIC environment -
guide
conducted RF
IEC 61000-4-6
3Vrms/Veff
From 150 KHz
to 80 MHz
3 Vrms
The Piuma UP system must be used
in shielded places with a minimum
RF shielding efficiency and, for each
cable that enters the shielded room,
with a minimum RF filter attenuation
Recommended separation distances
d= 1.167 square root symbol P
d= 1.167 square root symbol P
80MHz - 800 MHz
d= 2.333 square root symbol P
800MHz - 2.5 GHz
where P is the maximum output
of the transmitter in Watt (W)
according to the transmitter
manufacturer and “d” is the
recommended separation distance
in metres (m).
The resistances of the RF fixed
transmitter fields, as determined
from a detection site (NOTE 3) must
be lower than the level of conformity
of each frequency range. (NOTE 4)
Interference may occur in the
proximity of equipment marked with
this symbol:
Radiated RF
IEC 61000-4-3
3 V/m
From 80 MHz
to 2.5 GHz
3 V/m
NOTE: The maximum frequency range is applied at 80MHz and 800MHz
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is influenced
by absorption and reflection from structures, objects and people.
NOTE: The intensity of the field for fixed transmitters, such as base stations for radio-
telephones (mobile and cordless telephones) and terrestrial mobile radios, amateur radio
equipment, AM and FM radio transmitters and TV transmitters cannot be known precisely
in theory. To assess an electromagnetic environment caused by fixed RF transmitters, consider
performing an on-site electromagnetic inspection. If the field intensity measured outside the
shielded room, in the place where a Piuma UP system is being used, exceeds the applicable
level of conformity indicated above, normal functioning of the Piuma UP system must
be monitored. If abnormal functioning is observed, additional measures may be necessary, such
as changing the position of the Piuma UP system or using a shielded room with more efficient
RF shielding and greater filter attenuation
NOTE: Above the frequency range of 150KHz and up to 80MHz, the resistance field
should be under 3 V/m
Recommended separation distances between portable and mobile radio communications
equipment and the Piuma UP system
The Piuma UP system is intended to operate in an electromagnetic environment in which
radiated RF disturbances are under control. The client or operator of the Piuma UP system
can contribute to preventing electromagnetic interference by ensuring a minimum distance
between the mobile and portable RF communications devices (transmitters) and the Piuma
UP system as recommended below, in relation to the maximum output power of the radio
communication equipment.
For transmitters specified for a maximum output power not listed above, the recommended