Commissioning and configuration

48

RFM - HF RFID Series

Leuze electronic

You can use RF-Config to configure and parameterize all RFID devices from the Leuze 
electronic product line. The various device types with their parameter sets are available for
selection via a device list. 

After starting the program, first select the connected device from the left side of the user
interface in the device bar.

Its factory configuration (parameter set) is displayed on the right side of the user interface
in three tabs. 

Notice!

In the 

Help

 menu, you will find the documentation for the selected device type:

• Technical description
• Instruction set and configuration structure
• Commissioning tips
• Data sheet for the standard HF transponders available from Leuze 

Transponder tab 

In this tab, you can select from the available transponder types. The various types have
different memory sizes and memory areas; please take note. The basic setting is release
for ICODE 1 and ICODE SLI transponders.

Figure 7.2:

RF-Config – Transponder tab

In addition, the operating mode for multiple transponders in the field simultaneously can be
activated and serial number transmission in the telegram can be added. 

Содержание RFM 12 SL 200

Страница 1: ...RFM series the SMART solution Stationary RFID read write systems transponders T E C H N I C A L D E S C R I P T I O N en 01 2011 03 50115614...

Страница 2: ...4 4961888 Fax Int 38 044 4961818 C Illies Co Ltd Tel Int 81 3 3443 4143 Fax Int 81 3 3443 4118 Profa Tech Ltd Tel Int 254 20 828095 6 Fax Int 254 20 828129 Leuze electronic Co Ltd Tel Int 82 31 382822...

Страница 3: ...2 SL 200 Ex n 13 3 2 5 Special additional characteristics of the RFM 62 SL 200 14 3 2 6 Device construction 15 3 3 Read and write techniques 15 3 4 Standalone connection 18 3 5 Networking Leuze multiN...

Страница 4: ...reader using the RF Config software 47 7 5 Acknowledgments and error codes 51 7 6 LED indicators 52 7 7 Transponder memory organization 53 8 Diagnostics and troubleshooting 54 8 1 General causes of e...

Страница 5: ...ice This symbol indicates text passages containing important information 1 2 Declaration of conformity The devices of the RFM HF RFID 13 56MHz series including corresponding TFM transponders and MA co...

Страница 6: ...5693 3 RFID Radio Frequency IDentification umbrella term for the contactless identification of objects equipped with transponders by means of radio waves HF High Frequency radio frequency band in whic...

Страница 7: ...manufactured and tested in accordance with the applicable safety standards They correspond to the state of the art 2 3 Intended use Attention The protection of personnel and the device is guaranteed o...

Страница 8: ...ices for the RFM 32 SL 200 Ex n devices Safety notices for use in potentially explosive areas of Group II Category 3 Zones 2 Gas Ex and 22 Dust Ex Attention This product may only be used by qualified...

Страница 9: ...ns The applicable regulations for the installation of electrical equipment in potentially explosive areas must be observed Operation of the RFM 32 SL 200 Ex n in potentially explosive areas is recomme...

Страница 10: ...his operating manual are not authorized Safety regulations Observe the locally applicable legal regulations and the rules of the employer s liability insurance association Qualified personnel Mounting...

Страница 11: ...rious antenna sizes and operating ranges Figure 4 1 RFM device series In general the RFID devices of the RFM series are designed for handling and warehousing systems with small to medium operating ran...

Страница 12: ...ield ensures very reliable detection Large opening angle semi spherical shape thus a large reading field width Compact construction optimized for the operating range Reliable detection of the transpon...

Страница 13: ...or of the RFM 12 SL 200 Figure 3 1 Typical reading behavior RFM 12 read write unit Attention The values specified may deviate as a result of temperature influences installation site read angle etc Tra...

Страница 14: ...gure 3 2 Typical reading behavior RFM 32 read write unit Attention The values specified may deviate as a result of temperature influences installation site read angle etc Transponder TFM 02 Transponde...

Страница 15: ...4 transponders in field simultaneously Certified for use in potentially explosive areas in Zones 2 gas and 22 dust Typical reading behavior of the RFM 32 SL 200 Ex n Figure 3 3 Typical reading behavio...

Страница 16: ...200 Figure 3 4 Typical reading behavior RFM 62 read write unit Attention The values specified may deviate as a result of temperature influences installation site read angle etc Transponder TFM 02 Tran...

Страница 17: ...For exact data please refer to the specifications of the read write systems see also Typical reading behavior from page 11 The front of the device black or the active side equipped with LEDs depends...

Страница 18: ...ect recognition in handling and warehousing systems Commissioning systems in dispatch centers Flexible material flow control in assembly lines and daisy chained manufacturing cells Acquisition of oper...

Страница 19: ...ameter of 50mm whereby a spacer height of 10mm is sufficient for a range of approx 50mm For the read write system it is recommended that a space in front of and along the sides equal to the device dim...

Страница 20: ...m Networking is via a parallel connection of the individual RS 485 interfaces of the respective MA 21 100 2 units Leuze multiNet plus In Leuze s proprietary multiNet plus network the individual networ...

Страница 21: ...t This is used for example on a packet conveyor system on which the transponder can be affixed on the right side on the left side or on top thereby requiring three read stations So that the host does...

Страница 22: ...lect isolation device connection with terminals service interface in the MA for device configuration MA 21 110 50030482 RS232 gateway to Profibus DP device connection with system plug service interfac...

Страница 23: ...I Code 1 2 Texas Instruments Tag It Tag It HFI Infineon my d vicinity Electrical data Operating voltage VCC 12 30VDC Power consumption approx 0 9W 35mA typ at 24VDC Host interface RS 232 Baud rate 960...

Страница 24: ...nity Electrical data Operating voltage VCC 12 30VDC Power consumption max 1 2W typ 45mA at 24VDC Host interface RS 232 Baud rate 9600 Protocol 8 data bits 1 stop bit 1 start bit no parity Data frame S...

Страница 25: ...Infineon my d vicinity Electrical data Operating voltage VCC 12 30VDC Power consumption approx 2W 90mA typ at 24VDC Data interface RS 232 Baud rate 9600 Protocol 8 data bits 1 stop bit 1 start bit no...

Страница 26: ...yring card transponder PET PVC spacer Ultramid PPS Color self adhesive transponder white disc transponder black high temperature transponder brown black keyring card transponder white spacer black Tra...

Страница 27: ...e without quick heating 0 C 50 C 20 C 50 C 20 C 70 C 25 C 85 C 25 C 100 C 40 C 85 C 20 C 50 C 20 C 70 C 25 C 85 C 40 C 85 C 25 C 120 C 25 C 200 C up to 200 C up to 210 C up to 220 C up to 250 C TFM 05...

Страница 28: ...read approx 50ms block min 100 000 write cycles typical min 10 years of data retention Mechanical data Material PA6 PU Color black Weight 0 045kg TFM05 0 01kg TFM03 Dimensions see dimensioned drawing...

Страница 29: ...nic RFM HF RFID Series 27 TNT 35 7 24V 4 7 Dimensioned drawings Figure 4 1 Dimensioned drawing of the RFM 12 SL 200 read write system Figure 4 2 Dimensioned drawing of the RFM 32 SL 200 and RFM 32 SL...

Страница 30: ...s Leuze electronic Figure 4 3 Dimensioned drawing of the RFM 62 SL 200 read write system Figure 4 4 Dimensioned drawing of spacer mounting devices for transponders Spacer 85 HT Spacer Spacer 30 HT Spa...

Страница 31: ...ioned drawing of the TFM non Ex transponders TFM 03 1910 219 TFM 05 1910 219 TFM 02 2210 210 TFM 08 2125 220 TFM 04 1190 230 TFM 05 1110 210 TFM 05 2610 210 TFM 02 1125 220 TFM 05 2210 210 TFM 06 2225...

Страница 32: ...Specifications 30 RFM HF RFID Series Leuze electronic Figure 4 6 Dimensioned drawing of the TFM 0x 1x10 EX Ex transponder TFM 05 1110 EX TFM 05 1510 EX TFM 03 1110 EX...

Страница 33: ...sert or chapter 9 1 Save the original packaging for later storage or shipping If you have any questions concerning your shipment please contact your supplier or your local Leuze electronic sales offic...

Страница 34: ...ID Series Leuze electronic RFM 32 SL 200 and RFM 32 SL 200 Ex n by means of four fastening holes in the device with four screws RFM 32 fastening via four fastening holes Figure 5 2 RFU 32 fastening ho...

Страница 35: ...d corners or edges as damages could result Recommended distance to metal metal foils 5mm Intended for use in mechanically protected environments Disc transponder Screw connection of the disc transpond...

Страница 36: ...at a defined distance to metal The transponder temperature at the reading location must be within the operating temperature range The distance between two adjacent devices should be twice as large as...

Страница 37: ...s in motion Figure 5 4 Arrangement of the read write systems on the transport system 5 5 Cleaning Clean the housing of the RFM HF RFID Series with a soft cloth after mounting Remove all packaging rema...

Страница 38: ...lectrical connection with ready made connection cable 6 1 Safety notices for the electrical connection Attention Before connecting the device be sure that the supply voltage agrees with the value prin...

Страница 39: ...ons of the RFM read write systems 6 3 Cable lengths and shield Observe the following maximum cable lengths and shield types Table 6 1 Cable lengths and shield Colour PIN Signal gr GY ws WH Schirm SHIE...

Страница 40: ...ented in the standard Leuze data frame ASCII characters are transmitted Each connector unit makes available a separate RS 232 service interface that enables configuration of the RFM read write system...

Страница 41: ...nd RFM 32 devices are tuned so that the full operating range is ensured when mounted in the standard manner or in a mounting holder with no metal in the area around the active surface The RFM32 can al...

Страница 42: ...connector unit 9 pin Sub D connector simplifies the configuration To configure the PC with installed RF Config software administrator permissions not required is simply connected to the 9 pin Sub D pl...

Страница 43: ...ponder Trigger The device reads after a trigger signal has been supplied or after a software trigger command Switching output If the read is successful the device supplies a 300ms high pulse at the ou...

Страница 44: ...1m0d1d0t3t2t1t0Name CR LF Where y Year 2 m Month 2 d Day 2 t Tag number 4 and Name Device type e g RFM 32 Command R Executes a restart and resets the device to factory settings 0x52 Command syntax STX...

Страница 45: ...5 Tagtype is the transponder type see table 7 1 Data Data contents of the blocks specified in the command Notice If multitransponder mode anticollision is active serial number transmission must be act...

Страница 46: ...5 Notice When writing one or more complete blocks must always be transmitted i e 8 characters hex for 4 bytes block or 16 characters hex for 8 bytes block Command F Antenna electromagnetic field ON OF...

Страница 47: ...serial number of the transponder Block data With this setting 1 9 blocks can be read depending on transponder Answer telegram read 1 block beginning with block 05 e g 005014672616E Multiple pieces of...

Страница 48: ...es of information are contained in this answer starting at the front 0 Telegram flag only 1 telegram is output 05 Start block number 01 is the transponder type see table 7 1 4672616E is the data in he...

Страница 49: ...ntax STX Cyyzz CR LF where yy Register address zz Configuration data Answer STX Q1 CR LF see chapter 7 5 Notice The data is always hexadecimal encoded The number of pieces of data must be specified he...

Страница 50: ...erface in three tabs Notice In the Help menu you will find the documentation for the selected device type Technical description Instruction set and configuration structure Commissioning tips Data shee...

Страница 51: ...operation mode setting is important for automatic operation of the device Here you set the function after activation trigger operation mode and memory access block number Figure 7 3 RF Config Operati...

Страница 52: ...sive Listed below are some of the most important configuration constellations for which such dependencies exist If the Precharge function Active address 01 bit 6 Trigger must also Active address 01 bi...

Страница 53: ...pletely or with faulty characters Table 7 3 Possible error codes Notice If error code E08 occurs a CRC check was most likely accidentally activated To reset please send commands R and 0xD2 via the in...

Страница 54: ...No supply voltage READ Green continuous light and red continuous light for approx 3s Device OK initialization phase Self test running ACT PWR Yellow continuous light Device ready ACT Red continuous li...

Страница 55: ...hen writing Number of blocks 01 ICODE 44 05 11 4byte 02 ST_M LR512 60 00 16 4byte 03 TAG IT 32 00 08 4byte 04 ICODE SLI 112 00 28 4byte TFM 03 1110 210 50102917 TFM 05 1110 210 50102916 TFM 03 1510 21...

Страница 56: ...again Send device to customer service Status LED ACT red continuous light Device deactivated or device acti vated but no transponder detected Check wiring of trigger After activation and with transpon...

Страница 57: ...easures column which you have already examined fill out the following address field and fax the pages together with your service contract to the fax number listed below Customer data please complete L...

Страница 58: ...ry read write system for medium operating ranges up to 110mm for use in potentially explosive areas of Zones 2 22 50103087 Type designation Description Part no BT D30M 5 Mounting bracket for RFM 12 50...

Страница 59: ...transponder TFM 02 2210 210 50107790 IP 54 18x36x0 3mm 2g TFM 05 2210 210 50109232 IP 54 50x50x0 3mm 2g TFM 06 2225 220 50102913 IP 54 55x55x0 3mm 2g High temperature transponder TFM 05 2610 210 50109...

Страница 60: ...age the housing 10 2 Repairs and servicing Repairs to the device must only be carried out by the manufacturer Contact your Leuze distributor or service organization should repairs be required The addr...

Страница 61: ...NTROL 1 X ON Device control character 1 DC2 18 12 22 DEVICE CONTROL 2 TAPE Device control character 2 DC3 19 13 23 DEVICE CONTROL 3 X OFF Device control character 3 DC4 20 14 24 DEVICE CONTROL 4 Devic...

Страница 62: ...r than 63 3F 77 QUESTION MARK Question mark 64 40 100 COMMERCIAL AT Commercial AT A 65 41 101 A Uppercase letter B 66 42 102 B Uppercase letter C 67 43 103 C Uppercase letter D 68 44 104 D Uppercase l...

Страница 63: ...letter h 104 68 150 h Lowercase letter i 105 69 151 i Lowercase letter j 106 6A 152 j Lowercase letter k 107 6B 153 k Lowercase letter l 108 6C 154 l Lowercase letter m 109 6D 155 m Lowercase letter...

Страница 64: ...20 38 Dimensioned drawings 27 Disassembling 58 Disposing 58 E Electrical connection 36 Electromagnetic field 15 Error code 51 EtherCAT 20 38 EtherNet TCP IP 20 38 EtherNet IP 20 38 Ex transponder 26...

Страница 65: ...tective extra low voltage 37 Q Qualified personnel 8 Quality assurance system 3 R Read result 19 Repair 5 58 RF Config 9 47 RFID 4 RS232 20 RS422 20 38 RS485 20 38 S Safety notices 5 36 Service interf...

Отзывы: