Diagnostics and troubleshooting
Leuze electronic
DDLS 548i
59
8.2
Error displays and STATUS LED for remote diagnosis
Tab. 8.8:
STATUS LED displays - Causes and measures
LED
Color
State
possible causes
Measures
STATUS
LED
Green
Flashing
Warning message(s) set:
• SIGNAL QUALITY indicator
without green LED.
• Temperature, warning or error
(TMP).
• Laser pre-failure (LSR).
• Link Loss Counter has trig-
gered (LLC).
Check cause for the reduced SIG-
NAL QUALITY:
• Device alignment.
• Clean optical window.
• Eliminate the possibility of envi-
ronmental influences such as
snow, rain, fog.
• Laser diode: at end of life ex-
pectancy
Check LSR LED (see chapter
8.1 "Error displays of the operat-
ing state LEDs").
Check ambient temperature
• Initiate measures for lowering
the ambient temperature.
---
OFF
The transmitter is deactivated:
• No supply voltage.
• SIGNAL QUALITY indicator
shows only red LEDs.
• The LINK and LINK/ACT LEDs
are off.
Check supply voltage.
Check Ethernet cable connection.
Check cause for the reduced SIG-
NAL QUALITY:
• Device alignment
• Clean optical window
• Eliminate the possibility of envi-
ronmental influences such as
snow, rain, fog
• Laser diode: at end of life ex-
pectancy
Check LSR LED (see chapter
8.1 "Error displays of the operat-
ing state LEDs").
8.3
Error displays of the operating mode LEDs
Tab. 8.9:
ADJ LED displays - Causes and measures
LED
Color State
possible causes
Measures
ADJ
Gree
n
Flashing
• The "Adjust" operating mode is
not activated on the second
device.
• In the "Adjust" operating mode,
the supply voltage of the sec-
ond device was switched off/
interrupted.
Activate the "Adjust" operating mode on
the second device (see chapter 6.1 "Set-
ting the operating mode").