Starting up the device – Basic configuration
Leuze electronic
DCR 248i
45
TRIG
Trigger function that activates a read process with the configuration stored in the device, e.g., reading gate
control.
AUTO
By activating the
AUTO
function, the following sequence is started:
1. Optimum image setting: The device determines the optimum illumination setting for the given scenario.
2. Determine code types and number of digits: If codes are found, they are decoded.
3. Decoder table: The contents of the decoder table stored in the device are deleted. The new codes
(code type and number of digits) are stored in the decoder table.
NOTICE
Only activate the
AUTO
function while at a standstill!
Ä
Only activate the
AUTO
function if the code is not moving relative to the device.
NOTICE
AUTO
function not for Pharmacode!
Ä
The
AUTO
function cannot be used for Pharmacode codes.
ADJ
Adjustment function for aligning the device.
The reading quality is visually displayed as a percentage in the bar graph display. The bar graph display
depicts the average value over the last ten measurements.
NOTICE
Deactivate the
ADJ
function!
Ä
You must deactivate the
ADJ
function with the enter button
.
TEACH
With activation of the teach function, a present code is taught-in as a reference code.
During the teach event, the reading gate is opened and a code located in the read field is decoded. The de-
coded code is stored as a new reference code in the device.
NOTICE
Only one code in the read field during the teach function!
Ä
Upon activation of the teach function, only one code may be located in the read field of the
device.
NOTICE
TEACH
function not for Pharmacode!
Ä
The teach function cannot be used for Pharmacode codes.