Connection to Circuits
14
Power Analyzer NORMA 4000, NORMA 5000
EO1111G REV G
5.3.2 With
shunt
F
eh
ler! A
u
to
T
ext-E
in
trag
n
ich
t d
efin
iert.
Danger! Risk of electrocution! Do not touch sense terminals!
The sense terminals at the shunts are powered with the same
voltage as the power connections.
Shunts are not isolated.
Never touch the sense terminals at the shunts.
The connecting leads to the shunts should be as short as
possible in order to prevent noise voltages.
F
eh
ler! A
u
to
T
ext-E
in
trag
n
ich
t d
efin
iert.
Danger! Risk of electrocution!
Risk of injury when touching connections, internal circuits and
measuring devices that are not earthed.
Always adhere to the instructions regarding the sequence of
connection (see page 12).
1
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
2
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
3
U
I
N
L1
max. 10 m
LO
HI
Guard
*1
Last
ALL INPUTS MAX 1000V CATII TO
U
I
N
L1
max. 10 m
LO
HI
Guard
*1
Last
1
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
2
HI
CURRENT
LO
10 A max
PROBE
EXT.SHUNT
10 V max
HI
VOLTAGE
LO
1000 V max
3
ALL INPUTS MAX 1000V CATII TO
*1
We recommend using MCS measuring leads for triaxial shunts and MCP leads
for planar shunts. Triaxial shunts are equipped with guard connectors in the
plugs, while planar shuts are equipped with guard sockets.
All manuals and user guides at all-guides.com
all-guides.com