5
Maximum input 600VRMS / 1000A Pk
Maximum overload 1000V RMS / 10,000A
2.1.4 VA measurement (Single and 3 Phase)
(RMS and DC)
Measuring range ................
0-600kVA DC or 425kVA
in AC
Autorange Facility
40kVA, 400kVA,
600kVA
Resolution..........................
10VA in 40kVA
100VA in 400kVA
1kVA in 600kVA
Accuracy
VA > 2kVA ......
±
2.5% rdg
±
5 dgts
VA < 2kVA ......
±
0.08kVA
Frequency Range...............
DC and 30Hz to 1kHz
Current Range ...................
30A to 1000A Pk
Voltage Range ...................
40V to 600V RMS
Maximum input 600V RMS / 1000A Pk
Maximum Overload 1000V RMS / 10,000A
2.1.5 VAR measurement (Single and 3 Phase)
(RMS)
Measuring range ................
0-425kVAR
Autorange Facility
40kVAR, 400kVAR,
600kVAR
Resolution..........................
10VAR in 40kVAR
100VAR in 400kVAR
1kVAR in 600kVAR
Accuracy
VAR > 2kVAR .
±
2.5% rdg
±
5 dgts
VAR < 2kVAR .
±
0.08kVAR
Frequency Range...............
30Hz to 1kHz
Current Range ...................
30A to 1000A Pk
Voltage Range ...................
40V to 600V RMS
Maximum input 600V RMS / 1000A Pk
Maximum Overload 1000V RMS / 10,000A
2.1.6 Power Factor (Single and 3 Phase)
Measuring range 0.3 cap 1.0 ..0.3 ind
(72.5
ø
cap ..... 1.0 .... 72.5° ind)
Resolution..........................
0.01
Accuracy ............................
± 3°
Frequency range ................
30Hz to 1kHz
Voltage range.....................
40V to 600V RMS
Current range.....................
30A Pk to 1000A Pk
Measurement overload ......
1000V / 10,000A