Fig. 6

Current Measurement / Logging

4.  SAFETY

The instrument has been designed to comply with
IEC1010-2-032 Installation Category (Overvoltage
Category) III 600V Pollution degree 2 and
UL 3111-1. The product range conforms with the EEC
Low Voltage Directive 73/23/EEC and 93/68/EEC.
IEC 1010 is a safety standard which has the following
features:

 

Installation categories I to IV relate the maximum
working voltage to overvoltage transients that can
be expected in the measuring environment. For
the LH range of instruments, 600V CATIII, the
maximum expected transients must not exceed
6kV peak.

 

In a pollution degree 2  environment the internal
design of the instrument can cope with transient
conductivities due to condensation.

Safe operation of the instrument is the
responsibility of the operator who must be suitably
qualified and/or authorised.

Users of this equipment and or their employees
are reminded that Health and Safety Legislation
require them to carry out valid risk assessments of
all electrical work so as to identify potential
sources of electrical danger and risk of electrical
injury such as from inadvertent short circuits.

Содержание LH1050

Страница 1: ...LH1050 AC DC Clamp On Power Meter Operating Instructions...

Страница 2: ...l Data 4 2 2 General Data 6 3 OPERATING INSTRUCTIONS 7 3 1 Current Measurement 8 3 2 Voltage Measurement 9 3 3 Power 1 Measurement 10 3 4 Power 3 Measurement 11 3 5 Frequency Measurement 12 3 6 Loggin...

Страница 3: ...afety and electromagnetic compatibility European Low Voltage Directives 73 23 EEC and 93 68 EEC European EMC Directives 89 336 EEC and 93 68 EEC Submitted for approval to UL 3111 1 Safety Standards IE...

Страница 4: ...Push button switch for VA VAR and PF measurements in Watts mode Press and hold for 2 seconds for REC mode Min Max Av 5 Push button switch for HOLD saves complete data set when in single phase Watts mo...

Страница 5: ...2 1 2 Voltage measurement RMS and DC Measuring range 0 600V DC or AC Autorange facility 400V 600V Resolution 100mV in 400V range 1V in 600V range Accuracy V 40V 1 rdg 5 dgts V 40V 1V Maximum overload...

Страница 6: ...1000V RMS 10 000A 2 1 5 VAR measurement Single and 3 Phase RMS Measuring range 0 425kVAR Autorange Facility 40kVAR 400kVAR 600kVAR Resolution 10VAR in 40kVAR 100VAR in 400kVAR 1kVAR in 600kVAR Accurac...

Страница 7: ...ry for data logging The instrument outputs the displayed value Additional data is sent when in Watts and Hz settings Function Output Watts Watts VA VAR PF Hz Hertz Volts or Amps Output rate 2400 baud...

Страница 8: ...0 C 4 F to 140 F Maximum operating altitude 2000m 2 2 4 Mechanical Dimensions Length 251mm 9 88 Width 98mm 3 86 Depth 52mm 2 05 Weight 500g 1 1 lbs Case Material Bayblend T85MN Jaw Opening 55mm 2 2 Ac...

Страница 9: ...rotary switch to the Amps position Press the trigger to open the jaws and clamp them around the current carrying conductor as shown in Fig 2 Read the display Press the HOLD ZERO button to freeze the d...

Страница 10: ...ed value Press the HOLD ZERO button to freeze the display Press and hold the REC button 2 seconds to activate the REC mode Press the REC button again to cycle through Min Max Average readings Press an...

Страница 11: ...VAR VA PF REC button to cycle though the W VA VAR PF measurements In PF mode the bargraph displays the corresponding Watts value Press and hold for more than 2 seconds to enter or exit from the REC m...

Страница 12: ...negative lead to L3 Read the displayed value Use the VAR VA PF REC button to cycle though the W VA VAR PF measurements In PF mode the bargraph displays the corresponding Watts value Press and hold fo...

Страница 13: ...0 If ARMS 5 0A a volts frequency measurement will be made providing VRMS 5 0V otherwise will be displayed Press the HOLD ZERO button to freeze the display The bargraph displays the corresponding curre...

Страница 14: ...suring environment For the LH range of instruments 600V CATIII the maximum expected transients must not exceed 6kV peak In a pollution degree 2 environment the internal design of the instrument can co...

Страница 15: ...e unit is not hermetically sealed and should NOT be brought into contact with surface water Frequently inspect the test leads and the instrument for damage If the instrument is physically damaged or d...

Страница 16: ...hould be returned to us postage prepaid with a description of the defect The use of a battery other than that specified invalidates this warranty Goods alleged by the buyer to be defective shall not f...

Страница 17: ...ment solutions encompassing current measurement from 5mA to 2000A LEM HEME policy is one of continuous product improvement and the company reserves the right to revise the above specifications without...

Отзывы: