ITZ 2000-SBPR FLEX ULTRASTAB
System overview
Front panel indicators
When the mains supply voltage is set up, the rack electronics starts and initializes itself. After this process, the LEDs POWER,
STATUS and 2000 A light up.
The LEDs present on this panel display information about conditions which affect the operation of the transducer.
They are defined in the following table:
LED
Description
POWER (blue LED)
This LED is lit when the mains supply voltage is present
ACTIVE (yellow LED)
This LED is lit when
I
P
is higher than approximately 1 %
of
I
P N
DC
HIGH
I
P
(yellow LED)
This LED is lit when
I
P
is higher than approximately 105 %
of
I
P N
DC
OVERLOAD (red LED)
This LED is lit when the transducer head saturates due to
excessive primary current.
600 A ... 24000 A (yellow LED)
These LEDs are lit when a transducer head with the
corresponding full-scale range is connected to the rack
electronics. For ITZ 2000-SBPR, only LED 2000 A is lit, and
the others may be ignored.
CUSTOM (yellow LED)
Reserved for a custom head configuration.
STATUS (green LED)
This LED is lit when the unit status is OK (Normal operation).
POWER
2000A
16000A
OVERLOAD
600A
10000A
5000A
24000A
HIGH
I
P
ACTIVE
STATUS
CUSTOM
Front Panel Indicators
0.0000
U
in
IMPORTANT :
The GREEN secondary current path must be
closed at all times when MAINS power is
applied. Also, the programming plug must
always be mounted on the transducer head.
Mains
100-240
V
50/60
Hz
ITZ Rack Electronics
Power
Amplifier
I
s
.out
I
s
.return
0.1
Ω
Analog Out Connector
Monitor
Connector
±1
V full-scale
DMM
0.0000
I
in
DMM
Status/
Interlock
Connector
4
mm
Banana
Output
U
s
High
U
s
Low
R
M
I
s
.return
I
s
.return
Voltage Output Module
Transducer Head
Ferrite core
Programming
plug
Digital
Multimeter
Digital