8 PROCEDURE OF MEASUREMENTS
WARNING:
Use the instrument on voltage free systems only
Operation is easy :
Set measuring function with the central rotary switch
!
Connect instrument ! without measuring lead connected START is
omitted
!
Start measurement with "START" button
.
read out measured value.
For optimum performance and utilization of the device observe the following points:
8.1 General
8.1.1 POWER ON Functions
During switching on of the instrument with the central rotary switch it is possible to
access certain operating conditions by pressing certain button combinations:
a) Standard mode
If the device is put into operation without further button control, it switches into a
battery saving condition (Stand by-display "---") approx. 50 seconds after the
termination of a measurement, or after a button push or turn of the rotary switch.
Pressing the "display button" reactivates the instrument; the "old" measured values
can be read out again. After 50 min. of stand by the Display is turned off completely.
Instrument is reactivated with ON / OFF on the rotary switch
b) Stand by disable
A simultaneous push of buttons "DISPLAY" and "CHANGE
" during switching on
prevents the instrument from being switched off automatically (Stand by). The battery
saving mode is reactivated with ON / OFF on the central rotary switch.
c) Prolonged displaytest
By keeping the "DISPLAY" button pressed during switching on, the display test can
be prolonged for any length of time. Return to the standard operation mode by
pressing any button or turning of the central rotary switch.