6.5.2
Reaction on panic button
OFF
ON
PROG
*
.
9
.
7
.
1
.
3
1 7
#
.
#
.
*
.
Value
#
.
OFF
ON
PROG
Value Explanation
0 Silent
alarm
=>
listening-in connection
32
Un-delayed senior alarm =>
hands-free connection
(=factory setting).
6.6
Alarm criteria’s according to the position of the selection switch I/II/III
EasyAlarm®
monitors the emergency buttons on all three positions of the selection switch as well as the
optional attached motion detector checks for presence. But the acoustical monitoring of the room is deactivated.
(factory setting). This setting can be changed as follows:
A) Add up all the individual values according to your needs => Total=
n
.
Value
Explanation
1
Acoustical monitoring is locked on position III (high sensitivity)
2
Acoustical monitoring is locked on position II (medium sensitivity)
4
Acoustical monitoring is locked on position I
(low sensitivity)
8
Presence verification is locked on pos. III
16
Presence verification is locked on pos. II
32
Presence verification is locked on pos. I
*) If presence verification is locked, the motion sensor can be
activated as sensor-1-contact. In case of an activity (motion) an
alarm will be triggered (monitoring of the room or house
during your absence) => the total
n
must be supplemented with
the symbol '*'
see example below
Example:
Monitored activities
Selection switch on I:
Emergency presence control
Selection switch on II:
Emergency button + presence c acoustical monitoring
Selection switch on III: Emergency button + presence c acoustical monitoring
=> Acoustical monitoring locked on position I
=> Presence verification locked on position III => instead of presence verification an alarm shall be
triggered if motion is detected =>
n
= 4 + 8 + * = 12*
B) In case you would like to have an entry-delay for the sensor -1-contact, add following values:
Value v Explanation
0
Alarm due to sensor-1 without any delay
(=factory setting)
64
Alarm due to sensor-1 is delayed
Example:
Alarm due to sensor-1 shall be delayed =>
v
= 64
C) Program the total value (
s
=
v
+
n
= 64 + 12* = 76*) as follows:
1.
Slide
function switch
to PROG
2.
Enter
*
.
3.
Enter
2
.
6
.
8
.
4
.
<s>
(Example:
2
6
.
8
.
4
7
.
6
.
*
)
4.
Enter
#
.
5.
Enter
2
.
6
.
8
.
4
.
<s>
to
confirm (Example:
2
6
.
8
.
4
7
.
6
.
*
)
6.
Enter
#
.
¦
Correct => acknowledge beep
¦
Incorrect=> announcement „error“ => old value remains active.
7.
Slide
function switch
to OFF
Notes:
In case of incorrect programming, Slide
function switch
to OFF and repeat step 1 to 7.
Factory setting:
s
= 7, no acoustical monitoring; presence verification is activated.
6.7
How to program presence verification time-out
Time-out for presence verification is factory set to 24h. If there is no activation within this period, an alarm is
triggered after a pre warning period of 10 minutes (this period is indicated with a periodical loud beep tone).
OFF
ON
PROG
*
.
9
.
7
.
1
.
3
6 7
#
.
#
.
*
.
Value
#
.
OFF
ON
PROG
Value Explanation
0
Presence verification is switched off
6..255 Time-out in steps of 10 min. (min. 60min, max. 42.5h).
Factory setting
=144 (144*10min = 24 h)
Value