TS12P, Check & Adjust
51
Next step
6.
TOOLS T-Axis Adjustment Accuracy
<No.of Meas:>
Shows the number of runs executed. One run consists of a
measurement in face I and face II.
<
a T-axis:>
shows the standard deviation of the determined tilting axis
error. The standard deviation can be calculated from the second run
onwards.
It is recommended to measure at least two runs.
7.
MEAS
(F5) if more runs have to be added. Continue with step 3.
OR
CONT
(F1) to accept the measurements and to proceed to
TOOLS T-Axis
Adjustment Result
. No more runs can be added later.
Step
Description
IF the results are
THEN
to be stored
CONT
(F1) overwrites the old adjustment errors with the new
ones, if the
Use
status is set to
Yes
.
to be determined
again
REDO
(F2) rejects all new determined adjustment errors and
repeats the whole procedure. Refer to step 3. of paragraph
"Determination of tilting axis error step by step".