Preparation before Surgery
24
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01
7.10 Adjusting the tilt of the
microscope carrier (45°/70°)
1
X
X
Unscrew the cover (1).
X
X
Unscrew the 2 screws (2) and set the desired tilt.
2
X
X
Reattach and tighten the 2 screws (2).
X
X
Screw on the cover (1).
7.11 Balancing the Leica M525 F20/
CT20 surgical microscope
WARNING
Danger of fatal electrical shock!
X
X
The Leica M525 F20/CT20 surgical microscope may be
connected to a grounded safety socket only.
X
X
Connect the instrument to the power supply.
Mounting the accessories
WARNING
Risk of injury from downward movement of Leica M525 F20/
CT20 surgical microscope!
X
X
Never change the accessories or attempt to rebalance
the microscope while it is above the field of operation.
X
X
Always balance the Leica M525 F20/CT20 after
re-equipping it.
X
X
Do not exceed the max. load when equipping
components and accessories.
WARNING
Danger of tilting if load exceeds permitted load!
X
X
The load placed on the microscope interface must not
exceed 6.5 kg.
X
X
Mount the necessary accessories on the Leica M525 F20/CT20
surgical microscope (see chapter 9.7).
X
X
Switch on the power switch (2, chapter 6.2) on the control unit.
LED in power switch lights up.
Содержание M525 F20/CT20
Страница 1: ...MEDICAL DIVISION User Manual 10 715 156 Version 01 Leica M525 F20 CT20...
Страница 14: ...Safety Notes 12 Leica M525 F20 CT20 Ref 10 715 156 Version 01 Leica M525 CT20 ceiling mount 1 1 Nameplate...
Страница 60: ...58 Leica M525 F20 CT20 Ref 10 715 156 Version 01...
Страница 61: ...Leica M525 F20 CT20 Ref 10 715 156 Version 01 59...