Operation
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01 37
Adjust the brightness of the touch panel monitor:
X
X
Select "LCD Brightness" with the arrow keys (1).
The following menu window appears:
X
X
Press the "+" or "–" button briefly.
The brightness of the touch panel monitor is changed in increments
of 1.
X
X
Press and hold down the "+" or "–" button.
The brightness of the touch panel monitor is changed in increments
of 5.
8.9 Controlling the zoom and focus
WARNING
Severe tissue damage can result from an incorrect working
distance!
X
X
When using lasers, always set the working distance of
the microscope to laser distance and lock the microscope
in position.
WARNING
Failure of the zoom or focus motor can be dangerous for the
patient!
X
X
If the zoom or focus motor fails, set the zoom or focus
manually (see chapter 8.9).
You can adjust the zoom and focus using the "Magn."
(Magnification) and "WD" (Working Distance) adjustment bars in
the main menu or using a footswitch/handswitch.
Using the adjustment bars in the main menu
X
X
Press the "+" or "–" button briefly.
The zoom/focus value is changed in increments of 1.
X
X
Press and hold down the "+" or "–" button.
The zoom/focus value is changed in increments of 5.
Содержание M525 F20/CT20
Страница 1: ...MEDICAL DIVISION User Manual 10 715 156 Version 01 Leica M525 F20 CT20...
Страница 14: ...Safety Notes 12 Leica M525 F20 CT20 Ref 10 715 156 Version 01 Leica M525 CT20 ceiling mount 1 1 Nameplate...
Страница 60: ...58 Leica M525 F20 CT20 Ref 10 715 156 Version 01...
Страница 61: ...Leica M525 F20 CT20 Ref 10 715 156 Version 01 59...