Page 43 Operating Manual EM CPD300 12/18
Adjustments of Pressure Threshold:
Go to settings and set the desired empty pressure threshold (see table above) under
the “CO
2
bottle empty pressure” area.
Change CO
2
bottle empty pressure threshold value according to the list on page 45.
The CO
2
temperature can be estimated by measuring bottle surface with
thermometer. CO
2
temperature is then about 1-2 °C cooler then the bottle surface.
Press „ Update “ to confirm.