Page 5 Operating Manual EM CPD300 12/18
2.
Safety
Warning!
The EM CPD300 critical point dryer can be operated simply and safely, if the
system is operated according to the operating manual.
2.1
Safety concept of Leica Mikrosysteme GmbH
The use in accordance with regulations of the EM CPD300 critical point dryer
consists of critical point drying of biological or industrial samples with minimal
deformation of the specimens for subsequent SEM analysis.
Any other application without written approval of the manufacturer can lead to
damages and injuries of instrument and user. The manufacturer refuses to take any
responsibility for damage caused thereby.
The sample pressure chamber with a maximum working pressure of 150bar is
secured with a Software and Hardware Cut Off function.
Hardware controlled cut off
Bursting membrane at 105bar at 20 °C (+/-10%)
Software controlled cut off
Pressure at 80bar and temperature at 45°C
For safe operation it is absolutely necessary to set up and connect the device
according to the instructions in Chapter 3.
Содержание EM CPD300
Страница 1: ...From Eye to Insight OPERATING MANUAL EM CPD300 167180032 Version 12 18 ...
Страница 27: ...Page 24 Operating Manual EM CPD300 12 18 Press Exchange Cycles Set Exchange Cycles Example set to 12 ...
Страница 30: ...Page 27 Operating Manual EM CPD300 12 18 Press Start to activate program Test ...
Страница 37: ...Page 34 Operating Manual EM CPD300 12 18 and touch starting time to check finishing time 13 13 ...
Страница 48: ...Page 45 Operating Manual EM CPD300 12 18 After software update press back ...
Страница 52: ...Page 49 Operating Manual EM CPD300 12 18 Remove poral filter with spring from the nut ...