Page 16 Operating Manual EM CPD300 12/18
Programm Screen / Panel:
1 Activates key pad to enter program name.
2 Activated program is marked green.
3 Stirrer on / off with speed control.
4 Activation of auto version. If not highlighted manual version is active.
Only selectable in automated version.
5 Sets speed of CO
2
influx in pressure chamber.
Three possibilities: slow, medium, fast.
6 Switch to filler and holder panel. Display of filler and holder status.
7 Sets delay time after influx of CO
2
and before starting exchange process.
8 Sets exchange speed from 1-10.
The number relates to the setting of the needle valve during exchange. 1 means
a small opening and staying open for longer (slow flowing speed), 10 means a big
opening for a short amount of time (fast flowing speed). The parameters where
set in a way that after a minimum of 12 exchange cycles only liquid CO2 is left in
the chamber.
9 Sets exchange cycles. 12 cycles means one chamber volume is completely
exchanged. Minimum is 12 cycles.
10 Sets heating speed for critical point. Three possibilities: slow, medium, fast.
11 Sets gas out speed. Possibilities: slow, medium, fast.
4
3
2
13
5
6
7
8
9
10
11
12
1
Содержание EM CPD300
Страница 1: ...From Eye to Insight OPERATING MANUAL EM CPD300 167180032 Version 12 18 ...
Страница 27: ...Page 24 Operating Manual EM CPD300 12 18 Press Exchange Cycles Set Exchange Cycles Example set to 12 ...
Страница 30: ...Page 27 Operating Manual EM CPD300 12 18 Press Start to activate program Test ...
Страница 37: ...Page 34 Operating Manual EM CPD300 12 18 and touch starting time to check finishing time 13 13 ...
Страница 48: ...Page 45 Operating Manual EM CPD300 12 18 After software update press back ...
Страница 52: ...Page 49 Operating Manual EM CPD300 12 18 Remove poral filter with spring from the nut ...