106 z lamp housing, opened
1
Lid, flipped up,
2
Collector,
3
12 V/100 W halogen lamp or gas
discharge lamp (see Fig. 38),
4, 9
Lid mounting,
5
Reflector,
6, 8
Adjusting screws for x/y centring of reflector,
7
Focusing
of reflector,
10
Mounting screws for lamp holder,
11
Socket
for cut-out plug
Fig. 63
12 V/100 W lamp holder
• Unscrew the screw (59.1 or 61.1) on the lid
and remove the lid.
• Move the collector (61.3) to the front, if
required.
Note
This step is not required for the 107/2 lamp
housing.
Attention!
Leave the protective covering on the lamp
until the lamp is in its holder! Avoid making
finger prints or wipe them off immediately.
• Remove the defective lamp.
• Place a new 12 V/100 W halogen lamp straight
into the lamp holder (60.1 or 61.2).
• Move the collector back to its original
position.
• Replace the lid and secure it with the screw
(59.1 or 61.1).
• Attach the lamp housing to the microscope
and secure it with the lock screw.
• Connect the lamp housing to the power unit.
• Connect the microscope and the power unit to
the mains supply.
3
10
11
10
4
1
2
5
6
7
8
9
Содержание DMIL
Страница 1: ...Leica DMIL Instructions Bedienungsanleitung Mode d emploi...
Страница 3: ...3 Leica DM IL Instructions...
Страница 82: ...82 Notes Notes...
Страница 83: ...3 Leica DM IL Bedienungsanleitung...
Страница 162: ...82 Notizen...
Страница 163: ...3 Leica DM IL Mode d emploi...
Страница 242: ...82 Notes...
Страница 243: ......