Safety Notes
31
Leica HM500/Ref. 10 715 697/Version B
Dangers during use
Warning 1
Risk of eye injuries
➩
Do not look through the instrument optics while the
instrument self-test is in progress.
➩
Do not use the device until the instrument self-test is
completed.
Warning 2
Risk of eye injuries
• When lasers are used in combination with the
Leica HM500 microscope, it is the responsibility of the
user (laser safety officer) to ensure that eye protection
is sufficient and safe.
• For safety reasons, the protection class of the laser
safety glasses must be raised two classes (in accor-
dance with EN207) when standard eyepieces or inter-
changeable eyepieces I or II are used.
Caution 3
Danger of damage to the electronics
➩
Always connect the data cable first and then switch
on the instrument.
➩
Always switch off the instrument before disconnecting
the data cable.
➩
Never plug or unplug the data cable when the green
light in the charge level indicator is on.
Warning 4
Risk of eye injuries
➩
Before starting up the Leica MLS500 light source make
sure that nobody will be dazzled by the light-emitting
optics on the headset.
➩
Never look directly into the light outlets of the Leica
HM500 microscope, the Leica MLS500 light source or
the fiber-optic light guide when the instrument is in
operation.
➩
Never direct the light beam directly onto the eye.
➩
Never use the Leica MLS500 light source for eye
surgery or surgical procedures which require direct
illumination of the eye.
➩
With the Leica MLS500 light source switched on and
the fiber-optic light guide unplugged, reduce the
brightness control of the light source to the minimum
level.
Warning 5
Risk of burn injuries
➩
Never operate the fiber-optic light guide without
microscope optics.
➩
Never touch the fiber-optic light guide plug once it has
been unplugged from the Leica MLS500 light source.
Always let the plug cool down.
➩
Switch off the Leica MLS500 light source before
pulling out the fiber-optic light guide.
➩
After use, unplug the fiber-optic light guide from the
Leica MLS500 light source to prevent light accidentally
being emitted from the optical outlet end of the fiber-
optic light guide.
➩
Do not use a damaged or kinked fiber-optic light guide.
Warning 6
Risk to patients from failure of the Leica MLS500 light
source
➩
Change the lamp before the end of its theoretical
service life.
Warning 7
Risk of explosion
➩
Do not use the Leica MLS500 light source in areas
subject to explosion hazards.
➩
Do not use the Leica MLS500 light source with flam-
mable mixtures such as anesthetics, oxygen or disin-
fectants.
Caution 8
Risk of fire damage
➩
Keep a distance of at least 250 mm between the
microscope outlet optics and other objects.