Once range filtering is enabled the range filter icon
in the status bar is visible.
At system start the range filtering is disabled by default.
Available commands:
Field
Description
Min Range
Enable or disable range filtering for minimum range. All points
with a range lower than the entered limit will not be stored.
Max Range
Enable or disable range filtering for maximum range. All points
with a range higher than the entered limit will not be stored.
Command
Function
Store
Store current scan parameters and return to
Main Menu
.
MinDist
Open video camera window to select a point from video stream
image for a probe distance measurement which will be entered
in the
Min Range
field.
MaxDist
Open video camera window to select a point from video stream
image for a probe distance measurement which will be entered
in the
Max Range
field.
Page
Switch to the
Detail Scan
page.