65
Leica CM3600 XP
Operating the Leica CM3600 XP
7
•
Insert the specimen stage, non-orientable
and align by eye.
•
Attach the bases
and fasten the screws
sledge.
•
Check the specimen stage, non-orientable
for correct orientation and proper tightening.
6
Fig. 70
5
4
Fig. 71
7.6 Knife
Warning
Unintentional touching of the knife or unintentional movement of the microtome sledge
Severe cutting injuries and / or severe crushing injuries
•
Move the microtome sledge to the position which is most appropriate for your purpose and that
ensures the best safety.
•
Prior to any work involving the knife or the microtome, or inside the cryochamber, activate the
Emergency stop
switch, see
p. 18 – 3.3.1 Emergency stop switch)
•
For any work involving the knife or the microtome, or inside the cryochamber, wear the safety gloves
provided, see
p. 23 – 5.2 Standard delivery)
.
•
Prior to any work involving the knife or the microtome, or in the cryochamber, swing the knee lever
upwards, see
This chapter provides instructions on how to insert the knives.
•
Since there is a considerable risk of injury, the safety notes listed above must be observed.
•
Two types of knives can be used in the Leica CM3600 XP: