55
Leica CM3600 XP
Software
6
6.8.3 Error messages: Meaning and troubleshooting
Error code
Error message
Confirm elimination
Error 01
Knife holder – limit position
exceeded
•
Re-initialize instrument. If the problem recurs, call
Service Technician.
Error 02
Limit position of sledge
exceeded
•
Re-initialize instrument. If the problem recurs, call
Service Technician.
Error 03
Chamber – Alarm temperature
exceeded
•
Check whether there has been a power failure!
•
Check the set value of the alarm temperature.
It must differ at least 5 K from the chamber
temperature selected.
Error 04
Power failure
Error 05
Data transfer failed (happens
frequently subsequent to a
power failure)
•
Check the cable connection between PC and
control unit.
•
If the problem recurs, call Service Technician.
Error 06
• Not defined
Error 07
Emergency stop activated
•
Deactivate the emergency stop!
Error 08
Stepper motor – step error
•
Inform Service Technician.
Error 09
Refrigerating system – pressure
exceeded
•
Check the air circulation between the condenser
and the control unit.
•
Check the distance of the instrument to the wall (at
least 50 cm)!
•
If the problem recurs, call Service Technician.
Error 10
Flap limit switch malfunction
•
Re-initialize instrument. If the problem recurs, call
Service Technician.
Error 11
Call Service Technician
•
Call Service Technician.