Runt Trigger
The Runt trigger is programmed to occur when a pulse crosses a first threshold line and fails to cross a second
threshold line before recrossing the first. You can select both voltage thresholds within a time range of 100 ps to
20 s. Other defining conditions for this trigger are the edge (triggers on the slope opposite to that selected) and
runt width.
R
UNT
T
RIGGER
S
ETUP
1. Touch
the
Trigger
descriptor label.
2. If applicable, touch the
Smart
trigger button, then touch the Runt trigger button.
3. Touch inside the trigger
Source
field and select a source on which to trigger. If
4. you select an external trigger source, you will also need to set a coupling mode.
5. Touch
inside
the
Level
data entry field and enter a value, using the pop-up keypad.
6. Select Positive or
Negative
polarity.
7. Touch
the
Less Than
button and enter a pulse-width value in the
Upper Limit
field.
o
Or touch the
Greater Than
button and enter a pulse-width value in the
Lower Limit
field.
o
Or touch the
In Range
button. Then touch the
Delta
button to set up a nominal range, plus or minus a
delta value in seconds. Touch inside the
Nominal Width
and
Delta
fields and enter values, using the
pop-up numeric keypads. Alternatively, touch the
Limits
button to set up a precise pulse-width range
by setting
Lower Limit
and
Upper Limit
values, using the pop-up keypads.
o
Or touch the
Out Of Range
button and perform the same range setups as for
In Range
triggering.
Slew Rate Trigger
1. Touch
the
Trigger
descriptor label.
2. If applicable, touch the
Smart
trigger button, then touch the Slew Rate trigger button.
3. Touch inside the trigger
Source
field and select a source on which to trigger. If
4. you select an external trigger source, you will also need to set a coupling mode.
5. Touch
inside
the
Level
data entry field and enter a value, using the pop-up keypad.
6. Select
Positive
or
Negative
polarity.
7. Touch
the
Less Than
button and enter a pulse-width value in the
Upper Limit
field.
o
Or touch the
Greater Than
button and enter a pulse-width value in the
Lower Limit
field.
o
Or touch the
In Range
button. Then touch the
Delta
button to set up a nominal range, plus or minus a
delta value in seconds. Touch inside the
Nominal Width
and
Delta
fields and enter values, using the
pop-up numeric keypads. Alternatively, touch the
Limits
button to set up a precise pulse-width range
by setting
Lower Limit
and
Upper Limit
values, using the pop-up keypads.
o
Or touch the
Out Of Range
button and perform the same range setups as for In Range triggering.
DISPLAY FORMATS
1. In the menu bar, touch
Display;
then touch
Display Setup
in the drop-down menu.
WRXi-OM-E Rev C
71
2. Touch one of the
Grid
combination buttons.
Autogrid
automatically adds or deletes grids as you select more
or fewer waveforms to display.