![LeCroy WaveRunner 104MXI Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/lecroy/waverunner-104mxi/waverunner-104mxi_operators-manual_1866369121.webp)
O
PERATOR
’
S
M
ANUAL
WRXi-OM-E Rev B
121
Parameter
Description
Definition
Notes
Period
Period of a cyclic signal
measured as time between
every other pair of 50%
crossings. Starting with first
transition after left cursor,
period is measured for each
transition pair, with values
averaged to give final result.
Where: Mr is the number of
leading edges found, Mf the
number of trailing edges
found,
the time when
rising edge i crosses the x%
level, and
the time when
falling edge i crosses the x%
level.
Standard parameter.
Period@level
Period at a specified level and
slope for every cycle in
waveform.
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels
from noise in data.
Available with JTA2 and
XMAP options.
Phase
Phase difference between
signal analyzed and signal
used as reference. You can set
the output type to percent,
degrees, or radians. After
setting up the reference, touch
the
More
tab for signal setups.
Phase difference between
signal and reference
Standard parameter.
Pop@X
Population at bin for specified
horizontal coordinate. You can
place the cursor at any bin and
use either Absolute, Reference,
or Difference cursor shape.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
PW50
Average pulse width at the 50%
point between the local
baseline and the local peak or
trough.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
PW50-
Average pulse width at the 50%
point between the local
baseline and the local trough.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
PW50+
Average pulse width at the 50%
point between the local
baseline and the local peak.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.