![LeCroy WaveRunner 104MXI Скачать руководство пользователя страница 119](http://html1.mh-extra.com/html/lecroy/waverunner-104mxi/waverunner-104mxi_operators-manual_1866369119.webp)
O
PERATOR
’
S
M
ANUAL
WRXi-OM-E Rev B
119
Parameter
Description
Definition
Notes
Maximum
Measures highest point in
waveform. Unlike top, does not
assume waveform has two
levels.
Highest value in waveform
between cursors
Gives similar result when
applied to time domain
waveform or histogram of data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition. Computes
horizontal axis location of
rightmost non-zero bin of
histogram - not to be confused
with maxp.
Standard parameter.
Max populate
Peak with maximum population
in a histogram.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
Mean
Average of data for time
domain waveform. Computed
as centroid of distribution for a
histogram.
Average of data
Gives similar result when
applied to time domain
waveform or histogram of data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition.
Standard parameter.
Median
The average of base and top
values.
Average of Base and Top.
Standard parameter.
Minimum
Measures the lowest point in a
waveform. Unlike base, does
not assume waveform has two
levels.
Lowest value in waveform
between cursors
Gives similar result when
applied to time domain
waveform or histogram of data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition.
Standard parameter.
Mode
Position of the highest
histogram peak.
Available with DDM2, JTA2,
and XMAP options.
Standard in DDA-5005A.
Nb phase
Provides a measurement of the
phase at a specific frequency of
a waveform (narrow band).
Available with DDM2 and
XMAP options.
Standard in DDA-5005A.
Nb Power
Provides a measurement of the
power at a specific frequency of
a waveform (narrow band).
Available with DDM2 and
XMAP options.
Standard in DDA-5005A.