114
WRXi-OM-E Rev B
Parameter
Description
Definition
Notes
Edge@level
Number of edges in waveform.
Reference levels and edge-
transition polarity can be
selected. Hysteresis argument
used to discriminate levels from
noise in data.
Available with JTA2 and XMAP
options.
Excel
Performs measurements in Excel
by transferring one or two
waveforms and reading the
resulting parameter value.
Available with XMAP option.
Excel must be loaded on the
instrument.
Fall time
Fall time: Duration of falling edge
from 90-10%.
Threshold arguments specify two
vertical values on each edge used
to compute fall time. Formulas for
upper and lower values:
lower = lower thresh. x amp/100 +
base
upper = upper thresh. x amp/100 +
base
Time at upper threshold
minus Time at lower
threshold averaged
over each falling edge
On signals not having two major
levels (triangle or saw-tooth
waves, for example), top and
base can default to maximum
and minimum, giving, however,
less predictable results.
Standard parameter.
Fall 80-20%
Fall 80-20%: Duration of pulse
waveform's falling transition from
80% to 20%, averaged for all falling
transitions between the cursors.
Average duration of
falling
80-20% transition
On signals not having two major
levels (triangle or saw-tooth
waves, for example), top and
base can default to maximum
and minimum, giving, however,
less predictable results.
Standard parameter.
Fall@level
Fall at level: Duration of pulse
waveform's falling edges between
user-specified transition levels. See
also Rise@level.
Duration of falling edge
between transition
levels
On signals not having two major
levels (triangle or saw-tooth
waves, for example), top and
base can default to maximum
and minimum, giving, however,
less predictable results.
Standard parameter.
First
Indicates value of horizontal axis at
left cursor.
Horizontal axis value at
left cursor
Indicates location of left cursor.
Cursors are interchangeable: for
example, the left cursor may be
moved to the right of the right
cursor and first will give the
location of the cursor formerly on
the right, now on left.
Standard parameter.