4.
Installation
This section describes the instructions that you should abide when install the centrifuge to ensure your
safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
WARNING
:
●
Incorrect power supply may damage centrifuge.
●
Make sure the power source conforms to the required power supply
before connecting.
4.1
Location
(
1
)
Locate this centrifuge on a firm
、
flat and level table, ensure the four feet of this centrifuge
stand on the table firmly. Avoid installing on the slippery table-board that conveys vibration.
(
2
)
Ideal ambient temperature is
20℃±5℃
, avoid placing the centrifuge in direct sunlight if
temperature exceeds 30
℃
.
(
3
)
Keep clearances of at least 10cm on both sides of centrifuge and at least 30cm behind it to
ensure its cooling efficiency.
(
4
)
No heat or water resource nearby, otherwise easily cause sample temperature increase or
centrifuge failures.
4.2 Connection of the power cord and grounding
WARNING
:
●
Do not touch the power cord with wet hands to avoid electrical
shocks.
●
This centrifuge must be grounded properly.
(
1
)
This centrifuge is equipped with a 3P flat plug; grounding can be done just by plugging the
3P plug into the outlet.
(
2
)
A cover 10A outlet which has a good ground is required, and meet with local safety
requirements.