![lbx MCEN-15K-001 Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/lbx/mcen-15k-001/mcen-15k-001_user-manual_681703013.webp)
◼
Clockwise rotate parameter button
to increase parameter value; counter-clockwise
rotate parameter button
to decrease parameter value.
◼
You can quick set value by parameter button
, quickly rotate parameter button
to speed up parameter changing.
◼
There is a circulating function to increase/decrease parameters. Clockwise rotate
parameter button
, values circulate from small
→
large
→
maximum
→
minimum; counter-clockwise rotate parameter button
, values circulate from
large
→
small
→
minimum
→
maximum.
(
2
)
Set the time
◼
Press select button
, time value flash into time setting mode.
◼
Rotate parameter button
to set running time range from 30 seconds to 99
minutes.
◼
When time displays as HD, that showing long-running mode.
8.1.3 Start the operation
(
1
)
Press running button
to start running
◼
It will start timing once the rotor starts running, the screen displays remaining run time.
(
2
)
Inquiry and modify the operation parameters
◼
Press select button
, return to the preparation interface and displays setting
parameters. Press again the select button
, once the parameter symbol is flashing, rotate
parameter button
to modify parameters. Release button after 5 seconds, centrifuge will
return to normal operation mode and run process according to the new parameters.
◼
After time parameter has been modified, the operation time is not cleared and will
continue to be accumulated.
(
3
)
Alarm display
◼
If anything wrong in processing, the centrifuge will brake to stop automatically, and
display the error code on the time area. You can find out the malfunction causes by checking
the table 10-1, and make correct actions accordingly.
8.1.4 End the operation
(
1
)
The centrifuge will brake when it reaches the setting time or
button is pressed.
◼
When the rotor stops rotating, centrifuge will start beeping to alarm the operation was over.
(
2
)
Open the door
◼
The door can be released automatically when the operation was over.
◼
After door closed, you are able to press the
button to open it.
◼
After ending the operation, the program will store the setting parameters of this operation,
and will recall these parameters when restart the program.