10.2.2 In the case of power outage
If the door cannot be opened due to the power outage, open the door as the follows.
(
1
)
Make sure if the rotor is rotating.
◼
Listen carefully to ensure that no rotating sound can be heard.
(
2
)
Insert a door key into the hole to open the door lock.
◼
Identify the hole which is located on the top right side of the unit.
◼
Insert a door key into the hole and push forward to release the door lock, and then
open the door.
11. Instructions of rotor and tube
CAUTION
:
●
Read the instructions thoroughly, correct use rotor.
●
Do not exceed
the allowable maximum speed
of rotor
、
tube
and adapters
etc.,
be care that
the allowable maximum speed
of some adapters are lower than
the rotor
’
s.
11.1 The rotor instructions
11.1.1
Rotor structure
Figure11-1 the rotor profile
11.1.2 Available rotors and adapters
The rotor can be used as follows:
Locking screw nut Lid knob Lid
Rotor block Spindle hole Tube