User Manual
MCEN-15K-001
Please read the User Manual carefully before use and follow all operating and safety instructions
user manual
english, español
Страница 1: ...User Manual MCEN 15K 001 Please read the User Manual carefully before use and follow all operating and safety instructions user manual english español ...
Страница 2: ...ith the following information Serial number Description of the problem Your contact information Warranty This instrument is guaranteed to be free from defects in materials and workmanship under normal use and service for a period of 12 months from the date of invoice The warranty is extended only to the original purchaser It shall not apply to any product or parts that have been damaged due to imp...
Страница 3: ...er Warning notes indicate any condition or practice which if not strictly observed could result in personal injury or possible death CAUTION Possible damage to instrument Caution notes indicate any condition or practice which if not strictly observed or remedied could result in damage or destruction of the instrument NOTE Notes indicate an area or subject of special merit emphasizing either the pr...
Страница 4: ...aminated by solutions with toxic radioactive or pathogenic materials clean it according to the decontamination procedure that you are specified If you require services at site please sterilize and decontaminate it in advance and then notice the service center involved in the details of the particular materials Do not handle the power cord or turn on or off the POWER switch with wet hands to void e...
Страница 5: ...the rotor surface before using it Do not use the rotor if such abnormality found 2 Do not set the centrifuge speed override the allowable minimum speed of the rotor kits rotor or the other adapters Make sure to run it below the allowable minimum speed 3 Do not exceed the allowable imbalance 4 Use the rotor and tubes within their actual capacities 5 If the rotor is provided with a lid ensure it is ...
Страница 6: ... 61010 2 10 Construction in accordance with the following EMC standards EN 61326 1 FCC Part 15 Subpart B IECS 001 Associated EU guidelines EMC guidelines 2004 108 EC Instrument guidelines 2006 95 EC This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme à la norme NMB 001 du Canada 3 Required Operational Condition 3 1 Basic operational Conditions 1 Power 110V 240V 50Hz 60Hz ...
Страница 7: ...eys vibration 2 Ideal ambient temperature is 20 5 avoid placing the centrifuge in direct sunlight if temperature exceeds 30 3 Keep clearances of at least 10cm on both sides of centrifuge and at least 30cm behind it to ensure its cooling efficiency 4 No heat or water resource nearby otherwise easily cause sample temperature increase or centrifuge failures 4 2 Connection of the power cord and ground...
Страница 8: ...arts Figure 5 1 MC15K front view of centrifuge Figure 5 2 MC15K Rear view of centrifuge Power Outlet Power switch Air vents View Door Door lock hook Rotor Operation panel Door release hole LCD Display foot ...
Страница 9: ...r button Clockwise rotate to increase parameter values counter clockwise rotate to decrease parameter values Press down the button shift between speed and RCF display 4 Open lock button Press the button to open the door The button is not available when the centrifuge is running 5 Start Stop button Press down the button to start running The centrifuge will brake to stop running if press the button ...
Страница 10: ...ual capacity Do not exceed the actual capacity specified in the instruction manual 7 3 Keep the tubes balance See Table 7 1 for allowable imbalance of each rotor Although this centrifuge can accept sample balancing by eye we recommend that you use this centrifuge in a well balanced condition to prolong its life expectancy Never intentionally run the centrifuge under unbalanced condition even thoug...
Страница 11: ... it immediately and contact our service center Notify the alarm code if displayed 8 1 Normal Operation Turn on the POWER switch The centrifuge will display the running interface last time after passing the self checking process See figure 8 1 Figure 8 1 the last running interface Speed 15000rpm Running time 12 minutes The door lock is released 8 1 1 Rotor loading and removal CAUTION Load the rotor...
Страница 12: ...erclockwise CAUTION Check that the rotor is firmly tightened before each start 8 1 2 Set the operation parameters Press the button to select required parameters The parameter can be modified when the parameter is flashing Clockwise rotate the parameter button to increase parameter value counter clockwise rotate the parameter button to decrease parameter value Parameter button rotate faster paramet...
Страница 13: ...ration interface and displays setting parameters Press again the select button once the parameter symbol is flashing rotate parameter button to modify parameters Release button after 5 seconds centrifuge will return to normal operation mode and run process according to the new parameters After time parameter has been modified the operation time is not cleared and will continue to be accumulated 3 ...
Страница 14: ...t is in secured position and then close the door 2 The centrifuge gets into preparation mode and displays last running parameters 3 Press knob and hold speed up to reach the maximum speed While release knob during acceleration the centrifuge will start to decelerate and stop 9 Maintenance Do not follow the recommended instructions for cleaning or disinfecting may damage the centrifuge 1 Centrifuge...
Страница 15: ... the rotor rinse the rotor with water Apply a thin coat of silicon grease to the rotor when it is completely dried The rotor should be regular maintenance recommend to cleaning it each 3 months to ensure tube and rotor holes keep clean and then apply a thin coat of silicon grease 10 Troubleshooting 10 1 Frequent problems list This centrifuge is designed with self diagnosing function When something...
Страница 16: ...6 Set wrong speed The setting speed exceed the allowable range Modify the speed value E 10 86 Read the service manual Contact with service center Table 10 1 Frequent problems and solutions Alarm code E 1 9 are related to wrong operating You can continue running the centrifuge after the cause removed 10 2 How to open the door WARNING Never open the door while the centrifuge is running The door is o...
Страница 17: ...a door key into the hole and push forward to release the door lock and then open the door 11 Instructions of rotor and tube CAUTION Read the instructions thoroughly correct use rotor Do not exceed the allowable maximum speed of rotor tube and adapters etc be care that the allowable maximum speed of some adapters are lower than the rotor s 11 1 The rotor instructions 11 1 1 Rotor structure Figure11...
Страница 18: ...eed 2 0 g ml Sample density g ml 1 2 11 1 4 Autoclaving The rotor makes by high strength aluminum alloy material can be high pressure sterilization 121 1 0kg cm2 20 minutes CAUTION The lid of the rotor is made of plastics can not be high pressure sterilization only ordinary sterilization can be used 11 2 Tubes 11 2 1 Cleaning and sterilizing tubes please refer to the table below Table 11 2 Cleanin...
Страница 19: ...temperature is exceeded the tubes may be deformed When using a sterilizing chamber please operate as follows 1 Place tubes in vertical position mouths upward If tubes are placed sideways they may deform into an oval shape due to gravity 2 Remove screw nuts and inner covers to prevent from deformation or rupture 3 Wait until the sterilizing chamber cools down to the room temperature before the tube...
Страница 20: ...amage We do not guarantee this centrifuge and the rotor under the following conditions even if within the guarantee period expires 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by conveyance or relocation after installation 4 Failures caused by unauthorized disassembly or modification 5 Failures caused by using parts of the other comp...