16
screen below, the 6300-M4 has been commanded to achieve a climb of 3000 Feet/min. on the
Static output. A dive would be indicated by a negative value. The Climb “Target” value is
meaningful only when the Static output is in the “Control” mode. This is explained later in
this section. How to change the Climb “Target” value is explained in section 3.2.2.
Units: knots Feet/min
Feet
Actual: M 101.2
3500
1005 M
Target:
20.0 3000 0
Parameter 10. Pitot Mode:
The Pitot output can be in one of three modes –
Measure (“
M
”), Leak (“
L
”), or Control (“
C
”).
In the Measure mode, the pressure at the Pitot output is simply being measured and
displayed as the “Actual” value in the current Pitot units. The pressure is not being changed
by the 6300-M4. In this mode, the Pitot valve inside the 6300-M4 that is used to change the
pressure at the Pitot output is pneumatically isolated from the Pitot output. In the Measure
mode, the Pitot “Target” value is meaningless, since the 6300-M4 is not trying to achieve any
specific target value.
The Leak mode is pneumatically identical to the Measure mode. When either (Pitot or Static)
output is in Leak mode, the screen changes from the Main Operating screen to the Leak
screen. The Leak screen is detailed in section 3.3.
When the Pitot output is in the Control mode the 6300-M4 is actually attempting to achieve
the Pitot “Target” value. The Pitot control valve inside the 6300-M4 is now connected to the
Pitot output and is actually increasing or decreasing the pressure to achieve and maintain the
“Target” value.
For an explanation of how to change the Pitot mode between Measure, Leak and Control,
please refer to section 3.2.4.
Units: knots Feet/min
Feet
Actual: M 101.2
3500
1005 M
Target:
20.0
3000
0