Laurell Technologies
Quick Start WS-650 Series 10/2011 – 10070193.revL
All information contained in this manual is the property of Laurell Technologies Corporation® and is NOT to be edited, reproduced or
distributed without express written permission from a corporate officer.
- 8 -
Approximate Vacuum Achieved With
Different Pressure Settings
(3/8” fitting only)
Pressure (psi)
Vacuum in inches of Hg
10
1.3
15
3
20
4.4
25
6.2
30
8.5
35
11.3
40
14.8
45
18
50
24
55
25.3
60
25.3
NOTE: These values are approximations, actual values may vary.
NOTE: The processor has a minimum vacuum setting of 15”. This is a
factory setting. The processor must have 15” of vacuum to operate.