KEYSTROKES FOR SETUP
If the
MENU
key does not work, see Section 9 “Enabling & Locking Out Menu Items.”
Press Menu
Select Key
Press Digit
Select Key
Press Value Select
Key
_
InPut
Selection of signal
input type & range.
Needs to reflect
jumper settings.
_
AC U
AC Volts
__
0.2U
__
2.0U
_
20.0U
200.0U
600.0U
0.2, 2, 20, 200, 600V FS
_
AC A U
AC Amps
__
2.0a
_
20.0a
_
200.0a
_
5.0A
2, 20, 200 mA, 5A FS
SEtuP
Meter Setup
0
0_00
Display selection with
scale factor of 1
0
4-1/2 digits (±20,000)
1
5-digit remote display (±99,999)
2
4-1/2 digits, counts by 10 (±20,000)
3
3-1/2 digits (±2,000)
0
0
_00
Power line frequency
0
Noise minimized for 60 Hz
1
Noise minimized for 50 Hz
00_
0
0
0
Scale and offset method
1
Coordinates of 2 points method
2
Reading coordinates of 2 points method
00_0
0
Rear panel control
inputs 1 & 2
Connections
True
= logic 1 (0V or
tied to digital ground)
False
= logic 0 (5V or
open)
0
1 = Reset, 2 = Meter Hold
1
1 = Function Reset 2 = Pk or Valley Disp.
2
1 = Meter Hold
2 = Pk or Valley Disp.
3
1 = Meter Hold
2 = Tare
4
1 = Peak or Valley 2 = Tare
5
1 = Tare
2 = Reset
6
1 = 0, 2 = 0, decimal point 1= XXXXX
1 = 1, 2 = 0, decimal point 1 = XXXX
.
X
1 = 0, 2 = 1, decimal point 1 = XXX
.
XX
1 = 1, 2 = 1, decimal point 1 = XX
.
XXX
7
1 = 0, 2 = 0, decimal point 2 = XXXX
.
X
1 = 1, 2 = 0, decimal point 2 = XXX
.
XX
1 = 0, 2 = 1, decimal point 2 = XX
.
XXX
1 = 1, 2 = 1, decimal point 2 = X
.
XXXX
8
1 = Function Reset 2 = Display Blank
9
1 = Hold
2 = Display Blank
A
1 = Peak or Valley 2 = Display Blank
B
1 = Tare
2 = Display Blank
C
1 = Valley Display 2 = Peak Display
D
1 = Tare
2 = Tare Reset
Both inputs 1 and 2 set to 1 for selections
2
,
4
,
A
,
C
= Function Reset (fast, 20 msec).
Both inputs 1 and 2 set to 1 for selections
0
,
1
,
3
,
5
,
8
,
9
,
B
,
D
= Meter Reset (slow,
2 sec, same as removing & reapplying
power).