- 10 -
8. FRONT PANEL SETUP KEYS
Meter Front Panel
There are four front panel keys, which change function for the
Run Mode
and
Menu Mode
,
effectively becoming eight keys. The keys are labeled with alphanumeric captions (MENU,
PEAK, RESET, ALARMS) for the Run Mode and with symbols (
right arrow,
right triangle,
up triangle,
left arrow) for the Menu Mode.
FRONT PANEL LOCKOUT
The Menu Mode may not work with meters as received from the factory, because menu items
have been disabled in software. Disabling causes menu items not to appear in the menu
sequence. To enable menu items, change their lockout status under
Loc 1
,
Loc 2
and
Loc 3
from
“0” (locked or disabled) to “1” (unlocked or enabled), as explained in Section 9 of this
manual. Also ensure that the lockout jumper is not installed, as explained in that same section.
The paragraphs below assume that all menu items have been unlocked.
MENU MODE KEY ACTION
In the Menu Mode, pressing a key momentarily advances to the next menu item. Holding
down a key automatically advances through multiple menu items for fast menu navigation
.
KEYS IN RUN MODE
MENU Key.
Pressing
MENU
from the Run Mode enters the Menu Mode. Pressing
MENU
repeatedly will step the meter through the various menu items (if these have not been
locked out) and then back to the Run Mode.
PEAK Key.
Pressing
PEAK
normally causes the peak value of the input signal to be
displayed. The peak display then blinks to differentiate it from the normal present value
display. Pressing
PEAK
again returns the display to the present value. The
PEAK
key
can also be programmed to display Valley, alternating Peak or Valley, or to Tare the
reading to zero. When Peak or Valley is selected, periodic horizontals bars at the top of
the display indicate Peak, and periodic horizontals bars at the bottom indicate Valley.
RESET Key.
Pressing
RESET
with
PEAK
resets peak and valley values. Pressing
RESET
with
ALARMS
resets latched alarms. Pressing
RESET
with
MENU
performs a
meter reset (same as power on). Meter reset can also be applied via a rear panel connect
or a serial ASCII command.