- 26 -
RS232 CONNECTION
INSTRUMENT Connector
Pin
Signal
W100 TERMINAL
3 RS232:
TXD
4 RS232:
RXD
2 RS232:
SHIELD,
GND
COMMUNICATION SETTING
Into the
SERIAL COMMUNICATION SETTING
section (see receiving instrument manual), select
the desired serial port and operation mode: WEIRIP (
) or WEIMOD (
).
It’s not possible to enable this function on both serial ports; in case of conflict, the last
serial set, remains active.
Settable parameters:
-
: transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default: 9600).
-
: transmitting instrument address (only for
, from 1 to 99; default: 1).
-
: delay in milliseconds which elapses before the instrument replies (from 0 to 200 ms;
default: 0).
-
:
-
: no parity (default).
-
: even parity.
-
: odd parity.
-
:
stop bit (1 – 2; default: 1).
Receiving instrument parameters must be set with the same value of transmitting
instrument parameters.
Transmitting instrument
- SU
PPLY
RS232
- SU
PPLY
RS232
TX RX
RX
TX
Receiving instrument
Содержание W100
Страница 40: ......