background image

- 29 - 

SETPOINT PROGRAMMING 

 
From the weight display, press MENU to access the setpoint setting. 
 

MENU/ENTER:  to enter a menu/confirm the data entry. 

:        to modify the displayed figure or menu item. 

:        to select a new figure or modify the displayed menu item. 

ESC:        to cancel and return to the previous menu. 

 

 

 

-

 



: if the E/EC option is connected, it is possible to set 12 groups (classes) of different 

values for the setpoint; otherwise it is possible to set only the first class. Valid values for relays 
switching are selected by the E/EC position. 

 

-

 



 (from 0 to max full scale; default: 0): Setpoint; relay switching occurs when the weight 

exceed the value set in this parameter. The type of switching is settable (see section 

OUTPUTS 

AND INPUTS CONFIGURATION

). 

 

-

 



 (from 0 to max full scale; default: 0): Hysteresis, value to be subtracted from the setpoint 

to obtain contact switching for decreasing weight. For example with a setpoint at 100 and 
hysteresis at 10, the switching occurs at 90 for decreasing weight. 

 

These values are set to zero if the calibration is changed significantly (see sections 

THEORETICAL CALIBRATION 

and

 REAL CALIBRATION (WITH SAMPLE WEIGHTS)

). 

 
 









… 



… 





… 



Содержание W100

Страница 1: ...ENGLISH ENGLISH ENGLISH ENGLISH Installation and User Manual W100 W100ANA 2014 30 EU EN55022 2010 EN61000 6 2 2005 EN61000 6 4 2007 version 1 07 SYSTEM IDENTIFICATION...

Страница 2: ...er No guarantee against misuse Batteries Laumas provides 1 year guarantee from the date of delivery note against material defects or battery manufacturing faults Disposal of Waste Equipment by Users i...

Страница 3: ...WEIGHT ZERO SETTING 12 ZERO VALUE MANUAL ENTRY 12 REAL CALIBRATION WITH SAMPLE WEIGHTS 12 FILTER ON THE WEIGHT 13 ANTI PEAK 14 ZERO PARAMETERS 14 RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANGES 14...

Страница 4: ...TEST 28 DATE AND TIME SETTING 28 INFO MENU 28 SETPOINT PROGRAMMING 29 ALARMS 30 PRINTING EXAMPLES 32 RESERVED FOR THE INSTALLER 33 MENU LOCKING 33 MENU UNLOCKING 33 TEMPORARY MENU UNLOCKING 33 DATA D...

Страница 5: ...emote control switch coils Avoid inverters in the instrument panel if inevitable use special filters for the inverters and separate them with sheet metal partitions The panel installer must provide el...

Страница 6: ...l shifting and tipping relating to shocks and vibration windy conditions seismic conditions in the installation setting stability of the support structure EARTHING THE WEIGHED STRUCTURE By means of a...

Страница 7: ...itation wire must be equal or similar to the one indicated in the load cell data sheet input resistance The insulation value between the shield and any other cell wire and between any other cell wire...

Страница 8: ...o serial ports RS485 and RS232 for connection to PC PLC up to 32 instruments max 99 with line repeaters by ASCII Laumas or ModBus R T U protocol remote display printer The instrument can be connected...

Страница 9: ...F DECIMALS DISPLAY INCREMENTS 0 4 x 1 x 2 x 5 x 10 x 20 x 50 x 100 DIGITAL FILTER READINGS PER SECOND 0 012 7 s 5 300 Hz RELAY LOGIC OUTPUTS N 5 max 115 VAC 150 mA N 4 analog output version LOGIC INPU...

Страница 10: ...n 5 V max 24 V 23 LOAD CELL SIGNAL Sig 12 INPUT COMMON VDC 0 V 24 LOAD CELL SIGNAL Sig BASIC INFORMATION It is recommended that the power supply negative pole be grounded It is possible to supply up t...

Страница 11: ...2 23 IN 1 IN 2 IN COMMON INPUTS supply 5 24 VDC 0 20 4 20 mA COMMON 0 10 VDC 1 ANALOG OUTPUT OPTION 16 1415 4 WIRE load cell CONNECTION 24 SIGNAL 19 EXCITAT EXCITAT SIGNAL 21 20 23 22 12 EC OPTION 3 2...

Страница 12: ...function during weight display press and hold down MENU and keys at the same time press MENU immediately followed by KEY Short press Long press 3 s Into menus Semi automatic zero Tare resetting Cance...

Страница 13: ...9 MENU MAP Into menus changes are applied right after pressing the ENTER key no further confirmation is required SETPOINT SYSTEM PARAMETERS 3 s optional 3 s optional optional...

Страница 14: ...tion REAL CALIBRATION WITH SAMPLE WEIGHTS If the instrument HAS NOT BEEN CALIBRATED missing plant system identification tag proceed with calibration If load cells data are unknown follow the procedure...

Страница 15: ...ible to set a value between 0 50000 and 7 00000 mV V Example of 4 cell system with sensitivity 2 00100 2 00150 2 00200 2 00250 enter 2 00175 calculated as 2 00100 2 00150 2 00200 2 00250 4 The divisio...

Страница 16: ...g Confirming once again the weight is set to zero the value is stored to the permanent memory Press to display the value of the total weight reset by the instrument given by the sum of all of the prev...

Страница 17: ...le for more than 20 all the parameters with settable weight values are reset to default values LINEARISATION OPTION ON MAX 5 POINTS It is possible to perform a linearisation of the weight repeating th...

Страница 18: ...2 5 6 2500 12 5 7 4000 10 8 6000 10 9 7000 5 ANTI PEAK When the weight is stable the anti peak filter removes any sudden disturbances with a maximum duration of 1 second Confirm the filter on the weig...

Страница 19: ...mple if the parameter is set to 5 and is set to 2 the weight will be automatically set to zero for variations smaller than or equal to 10 x SETTING UNITS OF MEASURE These are the available units of me...

Страница 20: ...ue currently displayed pieces in set the weight of one piece newton metres the value set in will be multiplied by the weight value currently displayed kilogram metres the value set in will be multipli...

Страница 21: ...programmed setpoint value the contact will switch on the basis of weight according to setpoint see section SETPOINT PROGRAMMING the contact will not switch on the basis of weight but is controlled by...

Страница 22: ...d closing the input no operation is performed the input status may however be read remotely by way of the communication protocol closing the input for max one second the weight is transmitted over the...

Страница 23: ...g the tare value going back to the weight display the display shows the net weight subtracting the preset tare value and the NET LED lights up to show that a tare has been entered To delete a preset t...

Страница 24: ...nput closed the maximum weight value reached remains displayed By opening the input the current weight is displayed If you wish to use this input to view a sudden variation peak set the FILTER ON THE...

Страница 25: ...with the minimum analog setting the PLC or tester read 4 1 mA I must set the parameter to 3 9 to obtain 4 0 on the PLC or tester correction of analog output to full scale if necessary permit modifica...

Страница 26: ...he frequency set in item from 10 to 300 set set continuous weight transmission protocol to RIP5 20 60 RIP50SHA RIPLED series remote displays the remote display shows the net weight or gross weight acc...

Страница 27: ...40 Hz with minimum baud rate 4800 baud 80 Hz with minimum baud rate 9600 baud 100 Hz with minimum baud rate 19200 baud 200 Hz with minimum baud rate 38400 baud 300 Hz with minimum baud rate 38400 bau...

Страница 28: ...is necessary to connect the above mentioned resistors DIRECT CONNECTION BETWEEN RS485 AND RS232 WITHOUT CONVERTER Since a two wire RS485 output may be used directly on the RS 232 input of a PC or remo...

Страница 29: ...ection WEIMOD MODE see section WEIRIP MODE WARNING in order to use the weight reading via serial port the weight reading mode must be set as see section DATA DELETION AND PROGRAM SELECTION RS485 CONNE...

Страница 30: ...case of conflict the last serial set remains active Settable parameters transmission speed 2400 4800 9600 19200 38400 115200 default 9600 transmitting instrument address only for from 1 to 99 default...

Страница 31: ...y locked at power on and shows the instrument model To unlock it if necessary cut off the connection to the receiving instrument and follow the procedure in KEYPAD OR DISPLAY LOCKING section see trans...

Страница 32: ...n Test It allows the analog signal to range between the minimum and the maximum values starting from the minimum current output test voltage output test Millivolt Test displays the load cell response...

Страница 33: ...values for relays switching are selected by the E EC position from 0 to max full scale default 0 Setpoint relay switching occurs when the weight exceed the value set in this parameter The type of swit...

Страница 34: ...ble this message appears in the sample weight setting in real calibration after the fifth sample weight value has been entered the value set for the parameter is beyond the permitted values press ESC...

Страница 35: ...II __O F_ __O L_ __O F_ __O L_ __O F_ aa CR RIP __O F_ __O L_ __O F_ __O L_ __O F_ __O F_ HDRIP N _ERCEL _ER_OL _ER_AD _ER_OF O__SET CONTIN _ERCEL _ER_OL _ER_AD _ER_OF O__SET For RIP remote displays i...

Страница 36: ...r less than 3 seconds BASIC PRINTOUT W100 BASE Addr 01 DATE 12 09 11 14 48 12 GROSS 878 kg NET 589 kg TARE 289 kg BASIC PRINTOUT PEAK ENABLED W100 BASE Addr 01 DATE 12 09 11 14 48 12 GROSS 1204 kg NET...

Страница 37: ...to indicate that this menu item is unlocked TEMPORARY MENU UNLOCKING press and simultaneously for 3 seconds it is now possible to enter and modify all menus including those which are locked By return...

Страница 38: ...r approval indicating mandatory hardware code and serial number see section INSTRUMENT COMMISSIONING Configure the connection to the CLM serie intelligent junction box or to the multi channel weight t...

Страница 39: ...achstehenden Normen bereinstimmt F D claration de conformit Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s CZ...

Страница 40: ......

Отзывы: