Models S 1200, S 2400, S 4400, POU 3300, POU 3500
-
25
-
Problem
Possible Cause
Possible Solution
Repeated Liquid Level #1
alarms and the reservoir
needs filling
Leakage on the process
plumbing loop
LAUDA-Noah recommends using
an automatic leak detector to
locate Fluorinert™ leakages.
1.
Follow the operating
instructions in the leak
detector manual.
2.
Check along the process
plumbing, especially the
fitting and connectors.
3.
Repair any leaks. If leakage
problems persist, try using
two continuous hoses
between the module and the
process equipment.
Controller does not
communicate with host
equipment
1.
Controller is not in
Remote mode
2.
Communication
connector is not attached
securely
3.
Communication protocol
is not correct
4.
Controller address set
incorrectly
1.
Press the
Remote/Local
key
on the controller.
2.
Check communication
connectors.
3.
Select the appropriate
protocol in the Global Setup
Menu. See “Communications
Protocol” on page 4-26.
4.
CHX, CHX2 or RTU - Match
address to tool setting. RCM,
RCMT, or RCMe - Refer to
respective RCM manuals for
addressing scheme.