Flow controller MID 20
14
Figure
(not to scale)
Procedure when connecting the device components
1. Connect the outlet connector on the pressure side of the
Integral XT to the pressure input connector on the flow
controller (1).
2. Connect the outlet connector on the pressure side of the flow
controller to the input connector on the application (2).
3. Connect the outlet connector on the return of the application to
the input connector for the return on the flow meter (3).
4. Connect the outlet connector on the return side of the flow
controller (4) to the input connector for the return on the Integral
XT.
5. Connect the LiBus cable of the flow meter (5) to the LiBus
socket of the Integral XT.
6. Option when using the internal temperature sensor:
Connect the Lemo socket of the flow meter (6) with the Lemo
socket of the Integral XT. To do so, use a connecting cable.
This is available as an accessory
(
chapter 4.3).