Flow controller MID 20
22
7. Error messages for flow controller
If one of the following error messages appears on the display of the Integral XT, try to solve the problem
with the user action described.
Error messages
User action
Error 1305
Switch both devices off. Then wait approx. 20 s. Then, first switch on the
flow controller and then the Integral XT.
Error 1305 and
error 146
Switch both devices off. Then wait approx. 20 s. Then, first switch on the
flow controller and then the Integral XT.
Error 146
Switch off the Integral XT and check the LiBus cable connection between
the Integral XT and the flow controller respectively insert the LiBus cable.
Then, switch on the Integral XT.
If the error message persists, please contact the
LAUDA service department
(
chapter 8.2)
.